FIG 3. Aneurysmal pulsation measurements (± standard error of mean) obtained before and after coil embolization. (x axis, each time interval = 10–1 s)

A, After coiling and while circulating a 0.9% saline solution, a statistically significant (P = .022) reduction of wall displacement was observed

B, No statistically significant reduction was observed with blood.