AJDRAJNR - American Journal of Neuroradiology

This Article
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via CrossRef
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Barker, D. W.
Right arrow Articles by Lemley, T.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Barker, D. W.
Right arrow Articles by Lemley, T.

American Journal of Neuroradiology, Vol 14, Issue 3 587-590, Copyright © 1993 by American Society of Neuroradiology


ARTICLES

Balloon test occlusion of the internal carotid artery: change in stump pressure over 15 minutes and its correlation with xenon CT cerebral blood flow

DW Barker, CA Jungreis, JA Horton, S Pentheny and T Lemley
Department of Radiology, Presbyterian University Hospital, PA 15213.

PURPOSE: To determine whether stump pressure changes significantly over time during temporary internal carotid artery (ICA) occlusion via an endovascular balloon, and to examine correlations between stump pressure changes and cerebral blood flow (CBF) as measured by xenon CT. METHODS: Seventy candidates for ICA sacrifice were evaluated preoperatively with a test occlusion of the cervical ICA using an endovascular balloon. Measurements of ICA stump pressure above the occlusive balloon through a distal lumen in the balloon catheter were made throughout the 15-minute test. During occlusion, CBF was measured with stable xenon CT techniques. Patients were then categorized according to CBF results. RESULTS: The stump pressure changed significantly (P < .001) during the occlusive period. On average, stump pressure increased 7.7 mmHg (12%). Patients determined by xenon CT to have an asymmetric decrease in CBF with balloon occlusion did not display the upward trend in stump pressures and were significantly different (P = .013) in that respect from patients who demonstrated either no CBF change or only minimal bilaterally symmetric decreases in CBF. CONCLUSIONS: Large overlaps among the groups prohibit the use of stump pressure changes as an accurate predictor of CBF or of stroke risk in an individual patient. However, these data suggest an autoregulatory mechanism in the cerebral circulation that evolves over a period of minutes in contrast to the traditional concept of cerebral autoregulation which is thought to occur in a matter of seconds.


This article has been cited by other articles:


Home page
Am. J. Neuroradiol.Home page
N. Tomura, K. Omachi, S. Takahashi, I. Sakuma, T. Otani, J. Watarai, K. Ishikawa, H. Kinouchi, and K. Mizoi
Comparison of Technetium Tc 99m Hexamethylpropyleneamine Oxime Single-Photon Emission Tomograph with Stump Pressure During the Balloon Occlusion Test of the Internal Carotid Artery
AJNR Am. J. Neuroradiol., September 1, 2005; 26(8): 1937 - 1942.
[Abstract] [Full Text] [PDF]