Experimental arteriovenous fistulas: treatment with silicone-covered metallic stents
G Geremia, M Bakon, L Brennecke, M Haklin and B Silver
Department of Diagnostic Radiology, Rush-Presbyterian-St Luke's Medical Center, Chicago, Ill 60612, USA.
PURPOSE: To determine the efficacy of silicone-covered metallic stents in the treatment of experimentally created carotid-jugular fistulas. METHODS: Carotid-jugular fistulas were surgically constructed in six mongrel dogs. Silicone-coated, self-expanding metallic stents were placed across the fistula holes within the carotid artery, and carotid angiography was performed before, immediately after, and 4 and 8 weeks after stent placement. Fistula specimens were resected 2 months after stent placement and analyzed by means of gross and light microscopy. RESULTS: Angiography revealed complete closure of all fistulas immediately after stent deployment. The fistulas remained closed and all carotid arteries remained patent. Marked stenosis within the carotid lumen was seen along the proximal and distal ends of the stents. Gross and micropathologic specimens of the carotid-jugular fistulas revealed fibrous connective tissue and collagen across the fistula holes. Proliferative fibrous connective tissue, collagen, and fibromyoblasts were located at either end of the stents. The wires of the stents indented the intraluminal surface of the carotid arteries. CONCLUSIONS: Silicone-covered stents were effective in closing all experimentally created carotid-jugular fistulas. With further refinements and variations in technique, covered stents may prove a viable alternative to current endovascular devices.
This article has been cited by other articles:
![]() |
E. Archondakis, G. Pero, L. Valvassori, E. Boccardi, and G. Scialfa Angiographic Follow-Up of Traumatic Carotid Cavernous Fistulas Treated with Endovascular Stent Graft Placement AJNR Am. J. Neuroradiol., February 1, 2007; 28(2): 342 - 347. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Geremia, T. Brack, L. Brennecke, M. Haklin, and R. Falter Occlusion of Experimentally Created Fusiform Aneurysms with Porous Metallic Stents AJNR Am. J. Neuroradiol., April 1, 2000; 21(4): 739 - 745. [Abstract] [Full Text] |
||||