American Journal of Neuroradiology, Vol 19, Issue 5 917-920, Copyright © 1998 by American Society of Neuroradiology
ARTICLES |
Embolization of intracranial aneurysms with Guglielmi detachable coils augmented by microballoons
WP Sanders, TH Burke and BA Mehta
Department of Radiology, Henry Ford Hospital, Detroit, MI 48202, USA.
We describe two cases in which balloons were used to aid in the occlusion of intracranial aneurysms with Guglielmi detachable coils. In both patients, initial attempts to place coils within the aneurysmal sac failed, as the loops of the coil repeatedly protruded through a wide aneurysmal neck into the parent vessel. Temporary balloons placed across the neck of these small, wide-necked aneurysms, the bridging balloon technique, effectively narrowed the necks, providing support for the formation of a proper coil basket, which was not possible without the balloon. Follow-up studies at 1 year showed good results in both patients.
This article has been cited by other articles:
![]() |
K. De Keukeleire, P. Vanlangenhove, and L. Defreyne Evaluation of a Neck-Bridge Device to Assist Endovascular Treatment of Wide-Neck Aneurysms of the Anterior Circulation AJNR Am. J. Neuroradiol., January 1, 2008; 29(1): 73 - 78. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-P. Cottier, A. Pasco, S. Gallas, J. Gabrillargues, C. Cognard, J. Drouineau, L. Brunereau, and D. Herbreteau Utility of Balloon-assisted Guglielmi Detachable Coiling in the Treatment of 49 Cerebral Aneurysms: A Retrospective, Multicenter Study AJNR Am. J. Neuroradiol., February 1, 2001; 22(2): 345 - 351. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. S. Turk, A. H. Rappe, F. Villar, R. Virmani, C. M. Strother, and S. P. Lownie Evaluation of the TriSpan Neck Bridge Device for the Treatment of Wide-Necked Aneurysms : An Experimental Study in Canines Editorial Comment: An Experimental Study in Canines Stroke, February 1, 2001; 32(2): 492 - 497. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. M. Strother INTERVENTIONAL NEURORADIOLOGY AJNR Am. J. Neuroradiol., January 1, 2000; 21(1): 19 - 24. [Full Text] [PDF] |
||||
![]() |
T. J. Masaryk, J. Perl II, R. C. Wallace, M. Magdinec, and D. Chyatte Detachable Balloon Embolization: Concomitant Use of a Second Safety Balloon AJNR Am. J. Neuroradiol., June 1, 1999; 20(6): 1103 - 1106. [Abstract] [Full Text] |
||||