American Journal of Neuroradiology 20:1527-1534 (9 1999)
© 1999 American Society of Neuroradiology
ARTICLE
Detection of Intracranial Hemorrhage with Susceptibility-Weighted MR Sequences
a From the Department of Radiology, Kumamoto University School of Medicine, 1-1-1 Honjo, Kumamoto 860-8556, Japan.
BACKGROUND AND PURPOSE: Detection of hemorrhage is important in the diagnosis and management of a variety of intracranial diseases. We evaluated the sensitivity of the following sequences for depicting chronic hemorrhagic foci associated with susceptibility dephasing: gradient-recalled echo (GRE) imaging, GRE-type single-shot echo-planar imaging (GRE-EPI), spin-echotype single-shot echo-planar imaging (SE-EPI), turbo spin-echo (TSE) imaging, half-Fourier single-shot turbo spin-echo (HASTE) imaging, and segmented HASTE (s-HASTE) imaging. To our knowledge, no previous comparison has been made with these techniques in the same patient.
METHODS: Fifty patients with suspected chronic hemorrhage were examined prospectively with the above six sequences. Contrast-to-noise ratio (CNR), sensitivity to detection of lesions, conspicuity of internal architecture, and sensitivity to small hemorrhagic foci were evaluated.
RESULTS: Hemorrhagic foci were found in 35 patients. The CNR of the GRE, GRE-EPI, SE-EPI, TSE, s-HASTE, and HASTE sequences was 30.9, 23.7, 3.6, 6.1, -29.3, and -13.1, respectively; the number of small hemorrhagic foci detected was 85, 96, 44, 22, two, and one, respectively, for the supratentorial white matter; 70, 40, 19, four, zero, and zero, respectively, for the supratentorial cortical/subcortical region; and 73, 50, 26, 37, zero, and zero, respectively, for the infratentorial/skull-base region.
CONCLUSION: The GRE sequence was best for detecting susceptibility dephasing associated with chronic intracranial hemorrhage. GRE-EPI, while comparable to GRE in the supratentorial compartment, was reduced in its sensitivity near the skull base, and may be used as an alternative to GRE in uncooperative, unsedated, pediatric, or claustrophobic patients. SE-EPI should not be used in screening for intracranial hemorrhage.
This article has been cited by other articles:
![]() |
X. Lou, L. Ma, F.-L. Wang, Z.-P. Tang, H. Huang, Y.-Q. Cai, and E.-H.-C. Wong Susceptibility-Weighted Imaging in the Diagnosis of Early Basal Ganglia Germinoma AJNR Am. J. Neuroradiol., October 1, 2009; 30(9): 1694 - 1699. [Abstract] [Full Text] [PDF] |
||||
![]() |
S Sveinbjornsdottir, S Sigurdsson, T Aspelund, O Kjartansson, G Eiriksdottir, B Valtysdottir, O L Lopez, M A van Buchem, P V Jonsson, V Gudnason, et al. Cerebral microbleeds in the population based AGES-Reykjavik study: prevalence and location J. Neurol. Neurosurg. Psychiatry, September 1, 2008; 79(9): 1002 - 1006. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-M. Boulanger, S. B. Coutts, M. Eliasziw, A.J. Gagnon, J. E. Simon, S. Subramaniam, C.-H. Sohn, J. Scott, A. M. Demchuk, and for the VISION Study Group Cerebral Microhemorrhages Predict New Disabling or Fatal Strokes in Patients With Acute Ischemic Stroke or Transient Ischemic Attack Stroke, March 1, 2006; 37(3): 911 - 914. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Giugni, U. Sabatini, G. E. Hagberg, R. Formisano, and A. Castriota-Scanderbeg Fast Detection of Diffuse Axonal Damage in Severe Traumatic Brain Injury: Comparison of Gradient-Recalled Echo and Turbo Proton Echo-Planar Spectroscopic Imaging MRI Sequences AJNR Am. J. Neuroradiol., May 1, 2005; 26(5): 1140 - 1148. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Hermier and N. Nighoghossian Contribution of Susceptibility-Weighted Imaging to Acute Stroke Assessment Stroke, August 1, 2004; 35(8): 1989 - 1994. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. A. Walker, D. F. Broderick, A. L. Kotsenas, and F. A. Rubino Routine Use of Gradient-Echo MRI to Screen for Cerebral Amyloid Angiopathy in Elderly Patients Am. J. Roentgenol., June 1, 2004; 182(6): 1547 - 1550. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Tsushima, J. Aoki, and K. Endo Brain Microhemorrhages Detected on T2*-Weighted Gradient-Echo MR Images AJNR Am. J. Neuroradiol., January 1, 2003; 24(1): 88 - 96. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. D. M. Lin, C. G. Filippi, A. B. Steever, and R. D. Zimmerman Detection of Intracranial Hemorrhage: Comparison between Gradient-echo Images and b0 Images Obtained from Diffusion-weighted Echo-planar Sequences AJNR Am. J. Neuroradiol., August 1, 2001; 22(7): 1275 - 1281. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Munoz, N. J. Fischbein, J. de Vergas, J. Crespo, and J. Alvarez-Vincent Spontaneous Retropharyngeal Hematoma: Diagnosis by MR Imaging AJNR Am. J. Neuroradiol., June 1, 2001; 22(6): 1209 - 1211. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kinoshita, T. Okudera, H. Tamura, T. Ogawa, and J. Hatazawa Assessment of Lacunar Hemorrhage Associated With Hypertensive Stroke by Echo-Planar Gradient-Echo T2*-Weighted MRI Stroke, July 1, 2000; 31(7): 1646 - 1650. [Abstract] [Full Text] [PDF] |
||||