AJDRAJNR - American Journal of Neuroradiology

This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via CrossRef
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Shellock, F. G.
Right arrow Articles by Rasmussen, P. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Shellock, F. G.
Right arrow Articles by Rasmussen, P. A.

BRAIN

Aneurysm Clips: Evaluation of Magnetic Field Interactions and Translational Attraction by Use of "Long-Bore" and "Short-Bore" 3.0-T MR Imaging Systems

Frank G. Shellocka, Jean A. Tkachb, Paul M. Ruggierib, Thomas J. Masarykb and Peter A. Rasmussenc

a University of Southern California, Los Angeles, Keck School of Medicine and Institute for Magnetic Resonance, Safety, Education, and Research, Los Angeles, CA
b Division of Radiology, Neuroradiology Section, The Cleveland Clinic Foundation, Cleveland, OH
c Division of Surgery, Section of Neurosurgery, The Cleveland Clinic Foundation, Cleveland, OH

Address reprint requests to Frank G. Shellock, PhD, 7511 McConnell Avenue, Los Angeles, CA 90045

BACKGROUND AND PURPOSE: The use of 3.0-T MR systems is increasing worldwide. We evaluated magnetic field interactions and translational attraction for 32 aneurysm clips in association with exposure to "long-bore" and "short-bore" 3.0-T MR imaging systems.

METHODS: Thirty-two different aneurysm clips were evaluated in this investigation. Each aneurysm clip was qualitatively evaluated for magnetic field interactions and quantitatively assessed for translational attraction by using the deflection angle test. The deflection angle tests were performed at the points of the highest spatial gradients for long-bore and short-bore 3.0-T MR imaging systems.

RESULTS: Seventeen of the 32 aneurysm clips showed positive magnetic field interactions. Deflection angles for the aneurysm clips were significantly (P < .001) higher on the short-bore (range, 0–18 degrees) compared with those recorded on the long-bore (range, 0–16 degrees) 3.0-T MR imaging system. Aneurysm clips made from commercially pure titanium and titanium alloy displayed no translational attraction (n = 15), whereas those made from stainless steel alloy, Phynox, and Elgiloy displayed positive deflection angles (n = 17).

CONCLUSION: The 32 different aneurysm clips passed (angle <45 degrees) the deflection angle test by using the long- and short-bore 3.0-T MR imaging systems, indicating that they are safe for patients and other persons in these MR environments (ie, immediate area of MR imaging systems). However, only clips made from the titanium and titanium alloy are entirely safe for patients undergoing MR imaging procedures because of the total lack of magnetic field interactions. The remaining clips require characterization of magnetic field-induced torque. Because of possible differences in the points of the highest spatial gradients for different 3.0-T MR imaging systems, the results are specific to the imaging units and bore designs used in this investigation.




This article has been cited by other articles:


Home page
Am. J. Neuroradiol.Home page
H.-S. Kang, M. H. Han, B. J. Kwon, S. I. Jung, C.-W. Oh, D. H. Han, and K.-H. Chang
Postoperative 3D Angiography in Intracranial Aneurysms
AJNR Am. J. Neuroradiol., October 1, 2004; 25(9): 1463 - 1469.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Neuroradiol.Home page
A. Nehra, C. J. Moran, D. T. Cross III, and C. P. Derdeyn
MR Safety and Imaging of Neuroform Stents at 3T
AJNR Am. J. Neuroradiol., October 1, 2004; 25(9): 1476 - 1478.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
F. G. Shellock and J. V. Crues
MR Procedures: Biologic Effects, Safety, and Patient Care
Radiology, September 1, 2004; 232(3): 635 - 652.
[Abstract] [Full Text] [PDF]