American Journal of Neuroradiology 24:1819-1822, October 2003
© 2003 American Society of Neuroradiology
INTERVENTIONAL
Management of Neuroform Stent Dislodgement and Misplacement
From the Mallinckrodt Institute of Radiology, Washington University Medical Center, St. Louis, MO
Address correspondence to Christopher J. Moran, MD, Neuroradiology Section, Campus Box 8131, Mallinckrodt Institute of Radiology, Washington University School of Medicine, 510 S. Kingshighway Boulevard., St. Louis, MO 63110-1076
Summary: A self-expanding stent has recently been introduced for the treatment of wide-neck aneurysms. We describe two cases of stent malposition within large aneurysms. In the first case, the stent was dislodged during microcatheterization. This was managed by placement of a second stent through the interstices of the first followed by aneurysm coiling. In the second case, after deployment, the proximal portion of the stent moved into the aneurysm as the exchange guidewire was removed. This patient was treated by vessel sacrifice.
This article has been cited by other articles:
![]() |
S.H. Suh, B.M. Kim, T.-S. Chung, D.I. Kim, D.J. Kim, C.K. Hong, C.-H. Kim, J.Y. Ahn, and S.S. Kim Reconstructive Endovascular Treatment of Intracranial Fusiform Aneurysms: A 1-Stage Procedure with Stent and Balloon AJNR Am. J. Neuroradiol., January 1, 2010; 31(1): 155 - 160. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Yoo, D.J. Kim, D.I. Kim, J.-W. Lee, and S.H. Suh Bailout Stent Deployment during Coil Embolization of Intracranial Aneurysms AJNR Am. J. Neuroradiol., May 1, 2009; 30(5): 1028 - 1034. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Katsaridis, C. Papagiannaki, and C. Violaris Embolization of Acutely Ruptured and Unruptured Wide-Necked Cerebral Aneurysms Using the Neuroform2 Stent without Pretreatment with Antiplatelets: A Single Center Experience. AJNR Am. J. Neuroradiol., May 1, 2006; 27(5): 1123 - 1128. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.-W. Hsu, J.C. Chaloupka, J.A. Feekes, M.D. Cassell, and Y.-F. Cheng In vitro studies of the neuroform microstent using transparent human intracranial arteries. AJNR Am. J. Neuroradiol., May 1, 2006; 27(5): 1135 - 1139. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Pumar, M. Blanco, F. Vazquez, J. A. Castineira, L. Guimaraens, and A. Garcia-Allut Preliminary Experience with Leo Self-Expanding Stent for the Treatment of Intracranial Aneurysms AJNR Am. J. Neuroradiol., November 1, 2005; 26(10): 2573 - 2577. [Abstract] [Full Text] [PDF] |
||||
![]() |
O-K. Kwon, S. H. Kim, B. J. Kwon, H.-S. Kang, J. H. Kim, C. W. Oh, and M. H. Han Endovascular Treatment of Wide-Necked Aneurysms By Using Two Microcatheters: Techniques and Outcomes in 25 Patients AJNR Am. J. Neuroradiol., April 1, 2005; 26(4): 894 - 900. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. M. Pelz, S. P. Lownie, M. Kole, and D. Ramsay Desperate Appliance AJNR Am. J. Neuroradiol., January 1, 2005; 26(1): 198 - 200. [Full Text] [PDF] |
||||
![]() |
D. Pelz, P. Lylyk, and M. Negoro Interventional Neuroradiology Stroke, February 1, 2004; 35(2): 381 - 382. [Full Text] [PDF] |
||||