AJDRAJNR - American Journal of Neuroradiology

This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via CrossRef
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Niemann, D.
Right arrow Articles by Molyneux, A. J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Niemann, D.
Right arrow Articles by Molyneux, A. J.
Hotlight (NEW!)
Right arrow
What's Hotlight?

INTERVENTIONAL

Anatomically Conformable, Three-Dimensional, Detachable Platinum Microcoil System for the Treatment of Intracranial Aneurysms

David Niemannc,d, Richard Aviva,d, Caroline Cowsilla,d, Mary Sneadea,d and Andrew J. Molyneuxa,b,d

a Neurovascular Research Unit, Department of Neuroradiology, United Kingdom
b The Radcliffe Infirmary, United Kingdom
c Department of Neurological Surgery and Radiology, University of Wisconsin Medical School, Madison, WI
d Micrus Corporation, Mountain View, CA

Address reprint requests to Andrew J. Molyneux, MB, BChir, FRCR, Neurovascular Research Unit, Department of Neuroradiology, The Radcliffe Infirmary, Woodstock Road, Oxford OX2 6HE, United Kingdom

BACKGROUND AND PURPOSE: The configurations of detachable platinum coils have evolved over the last decade. Our objective was to assess the efficacy of the anatomically conformable three-dimensional (ACT) detachable platinum microcoil in the treatment of intracranial aneurysms.

METHODS: A series of 141 patients presented between May 2000 and August 2002 to a single neurointerventionalist for endovascular treatment of intracranial aneurysms. Eight patients were treated at another institute or are now overseas. The remaining 133 patients had 141 aneurysms, which were treated; 110 were ruptured and 31 were unruptured. The clinical conditions and aneurysm locations were recorded. The procedural technique was the same as that used with the Guglielmi detachable coil (GDC) device, though the coil detachment mechanism was unique. The ACT microcoil was used in almost all cases as the initial coil.

RESULTS: Complete (100%) occlusion was achieved in 104 (74%) aneurysms. Thirty four (24%) had subtotal occlusion (>=95%), and three (2%) had incomplete occlusion (<95%). The ACT coil conformed to both spherical and irregularly shaped aneurysms. No procedure-related deaths occurred, and there was no aneurysm rupture related to coil deployment. No postprocedural rebleeding was observed.

CONCLUSION: The ACT microcoil device is feasible as an alternative platinum coil device, offering some advantages over other currently available devices.




This article has been cited by other articles:


Home page
RadiologyHome page
O. N. Naggara, P. M. White, F. Guilbert, D. Roy, A. Weill, and J. Raymond
Endovascular Treatment of Intracranial Unruptured Aneurysms: Systematic Review and Meta-Analysis of the Literature on Safety and Efficacy