American Journal of Neuroradiology 25:1476-1478, October 2004
© 2004 American Society of Neuroradiology
INTERVENTIONAL
MR Safety and Imaging of Neuroform Stents at 3T
a Interventional Neuroradiology Service, Mallinckrodt Institute of Radiology
b Department of Neurological Surgery, Washington University School of Medicine, St. Louis, MO
c Department of Neurology, Washington University School of Medicine, St. Louis, MO
Address correspondence to Colin P. Derdeyn, MD, Mallinckrodt Institute of Radiology, 510 South Kingshighway Boulevard, St. Louis, MO 63110
Abstract
Summary: The Neuroform stent is a self-expanding nitinol stent designed for use in wide-necked intracranial aneurysms. Heating and imaging artifacts were evaluated by using a porcine carotid artery aneurysm model in a 3T MR system. A suspended Neuroform stent was tested for deflection. No heating was measured, and no evidence of deflection of the stent was found. Imaging artifacts were minimal. MR imaging in patients with Neuroform stent-treated aneurysms is safe and feasible.
This article has been cited by other articles:
![]() |
G. N. Levine, A. S. Gomes, A. E. Arai, D. A. Bluemke, S. D. Flamm, E. Kanal, W. J. Manning, E. T. Martin, J. M. Smith, N. Wilke, et al. Safety of Magnetic Resonance Imaging in Patients With Cardiovascular Devices: An American Heart Association Scientific Statement From the Committee on Diagnostic and Interventional Cardiac Catheterization, Council on Clinical Cardiology, and the Council on Cardiovascular Radiology and Intervention: Endorsed by the American College of Cardiology Foundation, the North American Society for Cardiac Imaging, and the Society for Cardiovascular Magnetic Resonance Circulation, December 11, 2007; 116(24): 2878 - 2891. [Abstract] [Full Text] [PDF] |
||||