American Journal of Neuroradiology 26:904-907, April 2005
© 2005 American Society of Neuroradiology
INTERVENTIONAL
Changes in the Intraaneurysmal Pressure Due to HydroCoil Embolization
a Department of Mechanical and Aerospace Engineering, University of California, San Diego, La Jolla, CA
b Department of Neurosurgery, Kaiser Permanente Medical Center, San Diego, CA
Address correspondence to David I. Levy, MD, Department of Mechanical and Aerospace Engineering, University of California, San Diego, 9500 Gilman Drive, MC 0411, La Jolla, CA 92093-0411
BACKGROUND AND PURPOSE: The purpose of this study was to measure the changes in the intraaneurysmal fluid pressure and parent vessel flow characteristics resulting from packing the aneurysmal sac with hydrogel-coated coils.
METHODS: Platinum coils coated with an expansible hydrogel were used to embolize a silicone model of the basilar tip aneurysm. The intraaneurysmal fluid pressure was measured with a NeuroCare Camino Pressure Micro-Probe at various packing ratios. A programmable pulsatile pump was used to input a physiologically relevant pulsatile flow in the parent artery.
RESULTS: The intraaneurysmal fluid pressure did not increase when packing the aneurysm with hydrogel-coated platinum coils, even with a coil density up to 93%.
CONCLUSION: Packing the aneurysm with hydrogel-coated coils at a density up to 93% did not increase the intraaneurysmal fluid pressure.
This article has been cited by other articles:
![]() |
T. Gunnarsson, F.C. Tong, P. Klurfan, C.M. Cawley, and J.E. Dion Angiographic and Clinical Outcomes in 200 Consecutive Patients with Cerebral Aneurysm Treated with Hydrogel-Coated Coils AJNR Am. J. Neuroradiol., October 1, 2009; 30(9): 1657 - 1664. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. K. Wakhloo, B. B. Lieber, G. Canton, D. I. Levy, J. C. Lasheras, and C. Geindreau Changes of Intra-Aneurysmal Pressure during Coiling AJNR Am. J. Neuroradiol., March 1, 2006; 27(3): 471 - 474. [Full Text] [PDF] |
||||