American Journal of Neuroradiology 27:822-829, April 2006
© 2006 American Society of Neuroradiology
BRAIN
Intracranial Time-Resolved Contrast-Enhanced MR Angiography at 3T
a Department of Biomedical Engineering, Northwestern University, Feinberg School of Medicine of Northwestern University, Chicago, Ill
b Department of Radiology, Feinberg School of Medicine of Northwestern University, Chicago, Ill
Address correspondence to Timothy J. Carroll, PhD, 676 N. St. Clair St, Suite 1400, Chicago, IL 60611
BACKGROUND AND PURPOSE: A method is presented for high-temporal-resolution MR angiography (MRA) using a combination of undersampling strategies and a high-field (3T) scanner. Currently, the evaluation of cerebrovascular disorders involving arteriovenous shunting or retrograde flow is accomplished with conventional radiographic digital subtraction angiography, because of its high spatial and temporal resolutions. Multiphase MRA could potentially provide the same diagnostic information noninvasively, though this is technically challenging because of the inherent trade-off between signal intensityto-noise ratio (S/N), spatial resolution, and temporal resolution in MR imaging.
METHODS: Numerical simulations addressed the choice of imaging parameters at 3T to maximize S/N and the data acquisition rate while staying within specific absorption rate limits. The increase in S/N at 3T was verified in vivo. An imaging protocol was developed with S/N, spatial resolution, and temporal resolution suitable for intracranial angiography. Partial Fourier imaging, parallel imaging, and the time-resolved echo-shared acquisition technique (TREAT) were all used to achieve sufficient undersampling.
RESULTS: In 40 volunteers and 10 patients exhibiting arteriovenous malformations or fistulas, intracranial time-resolved contrast-enhanced MRA with high acceleration at high field produced diagnostic-quality images suitable for assessment of pathologies involving arteriovenous shunting or retrograde flow. The technique provided spatial resolution of 1.1 x 1.1 x 2.5 mm and temporal resolution of 2.5 seconds/frame. The combination of several acceleration methods, each with modest acceleration, can provide a high overall acceleration without the artifacts of any one technique becoming too pronounced.
CONCLUSION: By taking advantage of the increased S/N provided by 3T magnets over conventional 1.5T magnets and converting this additional S/N into higher temporal resolution through acceleration strategies, intracranial time-resolved MRA becomes feasible.
This article has been cited by other articles:
![]() |
R. Habibi, M.M. Lell, R. Steiner, S.G. Ruehm, J.W. Sayre, K. Nael, and J.P. Finn High-Resolution 3T MR Angiography of the Carotid Arteries: Comparison of Manual and Semiautomated Quantification of Stenosis AJNR Am. J. Neuroradiol., January 1, 2009; 30(1): 46 - 52. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. M. Ruhl, M. Katoh, S. Langer, G. Mommertz, R. W. Guenther, T. Niendorf, and E. Spuentrup Time-Resolved 3D MR Angiography of the Foot at 3 T in Patients with Peripheral Arterial Disease Am. J. Roentgenol., June 1, 2008; 190(6): W360 - W364. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Y. Kim, R. A. Mirza, J. A. Bryant, E. D. Whiting, D. M. Delong, C. E. Spritzer, and E. M. Merkle Central Veins of the Chest: Evaluation with Time-resolved MR Angiography Radiology, May 1, 2008; 247(2): 558 - 566. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. K. Kuhl, F. Traber, and H. H. Schild Whole-Body High-Field-Strength (3.0-T) MR Imaging in Clinical Practice * Part I. Technical Considerations and Clinical Applications Radiology, March 1, 2008; 246(3): 675 - 696. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. R. Hadizadeh, M. von Falkenhausen, J. Gieseke, B. Meyer, H. Urbach, R. Hoogeveen, H. H. Schild, and W. A. Willinek Cerebral Arteriovenous Malformation: Spetzler-Martin Classification at Subsecond-Temporal-Resolution Four-dimensional MR Angiography Compared with That at DSA Radiology, December 1, 2007; 246(1): 205 - 213. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Wu, N. Kim, F.R. Korosec, A. Turk, H.A. Rowley, O. Wieben, C.A. Mistretta, and P.A. Turski 3D Time-Resolved Contrast-Enhanced Cerebrovascular MR Angiography with Subsecond Frame Update Times Using Radial k-Space Trajectories and Highly Constrained Projection Reconstruction AJNR Am. J. Neuroradiol., November 1, 2007; 28(10): 2001 - 2004. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Ali, T.A. Cashen, T.J. Carroll, E. McComb, M. Muzaffar, A. Shaibani, and M.T. Walker Time-Resolved Spinal MR Angiography: Initial Clinical Experience in the Evaluation of Spinal Arteriovenous Shunts AJNR Am. J. Neuroradiol., October 1, 2007; 28(9): 1806 - 1810. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Linn, B. Ertl-Wagner, K.C. Seelos, M. Strupp, M. Reiser, H. Bruckmann, and R. Bruning Diagnostic Value of Multidetector-Row CT Angiography in the Evaluation of Thrombosis of the Cerebral Venous Sinuses AJNR Am. J. Neuroradiol., May 1, 2007; 28(5): 946 - 952. [Abstract] [Full Text] [PDF] |
||||