AJDRAJNR - American Journal of Neuroradiology

This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via CrossRef
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Hatakeyama, Y.
Right arrow Articles by Korogi, Y.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hatakeyama, Y.
Right arrow Articles by Korogi, Y.

INTERVENTIONAL

Reduction of Radiation Dose for Cerebral Angiography Using Flat Panel Detector of Direct Conversion Type: A Vascular Phantom Study

Y. Hatakeyamaa, S. Kakedaa, N. Ohnaria, J. Moriyaa, N. Odaa, K. Nishinob, W. Miyamotob and Y. Korogia

a Department of Radiology, University of Occupational and Environmental Health School of Medicine, Kitakyushu, Japan
b Medical Systems Division, Shimadzu Corporation, Japan

Address correspondence to Shingo Kakeda, MD; Department of Radiology, University of Occupational and Environmental Health, 1-1 Iseigaoka, Yahatanishi-ku, Kitakyushu 807-8555, Japan; e-mail: kakeda{at}med.uoeh-u.ac.jp

BACKGROUND AND PURPOSE: Compared with image intensifier television (I.I.-TV) system, an angiography system using the flat panel detector (FPD) of direct conversion type has a high spatial resolution, which may improve image quality, reduce patient exposure, or both. Our purpose was to evaluate the detection of simulated aneurysmal blebs under dose reduction with the FPD system in comparison with the I.I.-TV system.

MATERIALS AND METHODS: A vascular phantom was designed to simulate various intracranial aneurysms with and without blebs, and this phantom was filled with 3 different concentrations of contrast material (300, 150, and 100 mg I/mL). 2D digital subtraction angiography (DSA) at low-dose mode of FPD system was compared with 2D DSA at a standard-dose mode of FPD system and a conventional mode of I.I.-TV system. Data analysis was based on 171 observations (57 aneurysms [20 with bleb and 37 without bleb] x 3 contrast material concentrations) by each of 7 radiologists, and the detection performances of blebs were compared using a receiver operating characteristic (ROC) analysis.

RESULTS: The mean dose measurements with a phantom during 2D DSA were 0.36 mGy/frame with low-dose mode of FPD system, 0.72 mGy/frame with standard-dose mode of FPD system and 0.76 mGy/frame with I.I.-TV system. The mean Az at 100 mg I/mL was significantly higher for low-dose mode of FPD than for conventional-dose mode of I.I.-TV mean Az, 0.85 versus 0.56; P < .01), though differences were not significant with 150 and 300 mg I/mL between both systems.

CONCLUSION: The FPD system allows a considerable dose reduction during 2D DSA without loss of the image quality.