AJDRAJNR - American Journal of Neuroradiology

This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via CrossRef
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Oouchi, H.
Right arrow Articles by Nishimura, T.
Right arrow Search for Related Content
PubMed
Right arrow Articles by Oouchi, H.
Right arrow Articles by Nishimura, T.

BRAIN

Diffusion Anisotropy Measurement of Brain White Matter Is Affected by Voxel Size: Underestimation Occurs in Areas with Crossing Fibers

H. Oouchia, K. Yamadaa, K. Sakaib, O. Kizua, T. Kubotaa, H. Itoa and T. Nishimuraa

a Department of Radiology, Graduate School of Medical Science, Kyoto Prefectural University of Medicine, Kyoto City, Kyoto, Japan
b Center for the Promotion of Excellence in Higher Education, Kyoto University, Kyoto City, Kyoto, Japan

Please address correspondence to Kei Yamada, MD, Department of Radiology, Kyoto Prefectural University of Medicine, Kajii-cyo, Kawaramachi Hirokoji Agaru, Kamigyo-ku, Kyoto City, Kyoto 602-8566, Japan; e-mail: kyamada{at}koto.kpu-m.ac.jp

BACKGROUND AND PURPOSE: Voxel size/shape of diffusion tensor imaging (DTI) may directly affect the measurement of fractional anisotropy (FA) in regions where there are crossing fibers. The purpose of this article was to investigate the effect of voxel size/shape on measured FA by using isotropic and nonisotropic voxels.

MATERIALS AND METHODS: Ten healthy adult volunteers had MR imaging by using a 1.5T clinical imager. DTI was performed with 2 different voxel sizes: a 2-mm-section isotropic voxel (2 x 2 x 2 mm3) and a 6-mm-section nonisotropic voxel (2 x 2 x 6 mm3). Images were obtained by using a single-shot echo-planar imaging technique with motion-probing gradients in 15 orientations and a b-value of 1000 s/mm2. FA and the apparent diffusion coefficient (ADC) were measured at different sites of the brain.

RESULTS: When smaller isotropic voxels were used, the FA was greater in areas with crossing fibers, including the superior longitudinal fasciculus, the thalamus, and the red nucleus; the FA was not significantly different in areas without crossing fibers, such as the corpus callosum, the posterior limb of the internal capsule, and the corticospinal tract at the level of the centrum semiovale (P > .05). The ADC values were not affected by voxel size/shape at any of the areas of the brain that were measured.

CONCLUSION: FA values that are measured in regions containing crossing fibers are underestimated when using nonisotropic DTI.




This article has been cited by other articles:


Home page
Am. J. Neuroradiol.Home page
S. Ito, W. Shirai, M. Asahina, and T. Hattori
Clinical and Brain MR Imaging Features Focusing on the Brain Stem and Cerebellum in Patients with Myoclonic Epilepsy with Ragged-Red Fibers due to Mitochondrial A8344G Mutation
AJNR Am. J. Neuroradiol., February 1, 2008; 29(2): 392 - 395.
[Abstract] [Full Text] [PDF]