BACKGROUND AND PURPOSE: Endovascular treatment of intracranial aneurysms using platinum coils is effective, but uncommonly aneurysms recur. New-generation coils, such as Cerecyte, aim to address this problem. This study examines the safety and efficacy of these coils in the treatment of a cohort of ruptured and unruptured aneurysms
MATERIALS AND METHODS: Sixty-seven patients with 68 aneurysms were included in the study. Of these, 51 were treated exclusively with the new polyglycolic acid (PGA)-containing coils, and 17 were treated with a combination of new PGA-containing and other coils. Initial and follow-up angiograms were graded according to the 3-point scale of occlusion. Follow-up angiography was available in 46 cases at 6 months. Based on occlusion grading at initial and follow-up angiography, aneurysms were classified into stable, improved, and worsened (recanalized) groups.
RESULTS: Of the exclusive new-coil cohort, 36 cases (70.6%) were initially completely occluded (grade 1), 12 (23.5%) showed filling at the neck (grade 2), and 3 (5.9%) showed contrast within the neck and sac (grade 3). Analysis of the follow-up angiograms showed 24 (70.6%) had stable occlusion, 3 (8.8%) had improved occlusion, and 7 (20.6%) had worsening occlusion. Data for cases treated with new PGA-containing coils together with bare platinum coils were also analyzed separately. Intraprocedural adverse events were noted in 4 cases (7.8%), but there were no clinical sequelae. There were no rebleeds in the follow-up period.
CONCLUSION: New PGA-containing coils show no excess in procedural and periprocedural complications over bare platinum coils, and the recanalization rate is comparable with bare platinum coils in the short term.
This article has been cited by other articles:
![]() |
M. Killer, D.F. Kallmes, M.R. McCoy, Y.-H. Ding, J.C. Shum, and G.M. Cruise Angiographic and Histologic Comparison of Experimental Aneurysms Embolized with Hydrogel Filaments AJNR Am. J. Neuroradiol., September 1, 2009; 30(8): 1488 - 1495. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Linfante, M.J. DeLeo III, M.J. Gounis, C.S. Brooks, and A.K. Wakhloo Cerecyte versus Platinum Coils in the Treatment of Intracranial Aneurysms: Packing Attenuation and Clinical and Angiographic Midterm Results AJNR Am. J. Neuroradiol., September 1, 2009; 30(8): 1496 - 1501. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. P. Ferns, M. E.S. Sprengers, W. J. van Rooij, G. J.E. Rinkel, J. C. van Rijn, S. Bipat, M. Sluzewski, and C. B.L.M. Majoie Coiling of Intracranial Aneurysms: A Systematic Review on Initial Occlusion and Reopening and Retreatment Rates Stroke, August 1, 2009; 40(8): e523 - e529. [Abstract] [Full Text] [PDF] |
||||
![]() |
P.M. White and J. Raymond Endovascular Coiling of Cerebral Aneurysms Using "Bioactive" or Coated-Coil Technologies: A Systematic Review of the Literature AJNR Am. J. Neuroradiol., February 1, 2009; 30(2): 219 - 226. [Abstract] [Full Text] [PDF] |
||||
![]() |
C.T. Whitlow, C.P. Geer, C.W.T. Mattern, B.J. Mussat-Whitlow, S.K. Yazdani, J.L. Berry, J.H. Lalli, R.O. Claus, V.R. Challa, and P.P. Morris Endovascular Histologic Effects of Ultrathin Gold- or Vitronectin-Coated Platinum Aneurysm Coils in a Rodent Arterial Occlusion Model: A Preliminary Investigation AJNR Am. J. Neuroradiol., January 1, 2009; 30(1): 85 - 90. [Abstract] [Full Text] [PDF] |
||||
![]() |
W.J. van Rooij, A.N. de Gast, and M. Sluzewski Results of 101 Aneurysms Treated with Polyglycolic/Polylactic Acid Microfilament Nexus Coils Compared with Historical Controls Treated with Standard Coils AJNR Am. J. Neuroradiol., May 1, 2008; 29(5): 991 - 996. [Abstract] [Full Text] [PDF] |
||||