AJDRAJNR - American Journal of Neuroradiology

Published ahead of print on August 21, 2008
doi: 10.3174/ajnr.A1235

This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Supplemental Online Table
Right arrow All Versions of this Article:
ajnr.A1235v1
29/10/1910    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via CrossRef
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Moritake, T.
Right arrow Articles by Tsuboi, K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Moritake, T.
Right arrow Articles by Tsuboi, K.

INTERVENTIONAL

Dose Measurement on Both Patients and Operators during Neurointerventional Procedures Using Photoluminescence Glass Dosimeters

T. Moritakea, Y. Matsumarub, T. Takigawab, K. Nishizawac, A. Matsumurad and K. Tsuboie

a RadGenomics Research Group, Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan
b Department of Endovascular Neurosurgery, Toranomon Hospital, Tokyo, Japan
c Radiological Protection Section, Research Center for Charged Particle Therapy, National Institute of Radiological Sciences, Chiba, Japan
d Department of Neurological Surgery, Doctoral Programs in Functional and Regulatory Medical Sciences, Graduate School of Comprehensive Human Sciences, University of Tsukuba, Ibaraki, Japan
e Proton Medical Research Center, University of Tsukuba, Ibaraki, Japan

Please address correspondence to Koji Tsuboi, MD, PhD, Proton Medical Research Center, University of Tsukuba, 1-1-1 Tennohdai, Tsukuba, Ibaraki, 305-8575, Japan; e-mail: tsuboi{at}pmrc.tsukuba.ac.jp

BACKGROUND AND PURPOSE: Although radiation skin injuries associated with interventional radiology have been known as a critical issue, there are few reports mentioning direct measurement of the entrance skin dose (ESD). Thus, the purpose of this study was to clarify the regional distributions of ESDs in neurointervention.

MATERIALS AND METHODS: Using photoluminescence glass dosimeters (PLDs), we measured the ESDs in 32 patients with a median age of 61.5 years. Angiographic parameters, including exposure time, dose-area product (DAP), and the number of digital subtraction angiography (DSA) studies and frames, were recorded. The ESDs of operators were analyzed by the same method.

RESULTS: The maximum ESD of 28 therapeutic procedures was 1.8 ± 1.3 Gy. Although the averaged ESD on the right temporo-occipital region was higher than that in other regions, disease-specific patterns were not observed. Statistically positive correlations were found between the maximum ESD and exposure time (r = 0.5283, P = .005), DAP (r = 0.7917, P < .001), the number of DSA studies (r = 0.5636, P = .002), and the number of DSA frames (r = 0.8583, P < .001). As for operators, ESDs to the left upper extremity were significantly higher than those to other regions. However, most of the ESDs were <0.2 mGy. Lead protective garments reduced the exposure doses to approximately one half to one tenth.

CONCLUSION: It was shown that the regional ESD could be measured by applying the PLD. This method should contribute to reducing the dose accumulation in patients as well as in operators.