AJDRAJNR - American Journal of Neuroradiology

Published ahead of print on April 10, 2008
doi: 10.3174/ajnr.A1073

This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
ajnr.A1073v1
29/7/1288    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via CrossRef
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Yang, C.-Y.
Right arrow Articles by Liu, H.-M.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Yang, C.-Y.
Right arrow Articles by Liu, H.-M.

BRAIN

Multiphase CT Angiography versus Single-Phase CT Angiography: Comparison of Image Quality and Radiation Dose

C.-Y. Yanga, Y.-F. Chena, C.-W. Leea, A. Huanga, Y. Shenb, C. Weib and H.-M. Liua

a Department of Medical Imaging, National Taiwan University Hospital and National Taiwan University College of Medicine, Taipei, Taiwan
b General Electrics Yokogawa Medical System, Tokyo, Japan

Please address correspondence to Hon-Man Liu, MD, Department of Medical Imaging, National Taiwan University Hospital and National Taiwan University College of Medicine, 7 Chung-Shan South Rd, Taipei, Taiwan; e-mail: hmliu{at}ntu.edu.tw

BACKGROUND AND PURPOSE: Conventional CT angiography (CTA) is acquired during only a short interval in the arterial phase, which limits its ability to evaluate the cerebral circulation. Our aim was to compare the image quality and radiation dose of conventional single-phase CTA (SP-CTA) with a multiphase CTA (MP-CTA) algorithm reconstructed from a perfusion CT (PCT) dataset.

MATERIALS AND METHODS: Fifty consecutive patients undergoing head CTA and PCT in 1 examination were enrolled. The PCT dataset was obtained with 40.0-mm-detector coverage, 5.0-mm axial thickness, 80 kilovolt peak (kVp), 180 mA, and 30 mL of contrast medium. MP-CTA was reconstructed from the same PCT dataset with an axial thickness of 0.625 mm by using a new axial reconstruction algorithm. A conventional SP-CTA dataset was obtained with 0.625-mm axial thickness, 120 kVp, 350 mA, and 60 mL of contrast medium. We compared image quality, vascular enhancement, and radiation dose.

RESULTS: SP-CTA and MP-CTA of 50 patients (male/female ratio, 31/19; mean age, 59.25 years) were analyzed. MP-CTA was significantly better than SP-CTA in vascular enhancement (P = .002), in the absence of venous contamination (P = .006), and was significantly higher in image noise (P < .001). MP-CTA used less contrast medium than SP-CTA and could demonstrate hemodynamic information. The effective dose of MP-CTA was 5.73 mSv, which was equal to that in conventional PCT, and it was 3.57 mSv in SP-CTA.

CONCLUSION: It is feasible that MP-CTA may provide both CTA and PCT results. Compared with SP-CTA, MP-CTA provides comparable image quality, better vascular enhancement, hemodynamic information, and more noise with less detail visibility with a lower tube voltage. The radiation dose of MP-CTA is higher than that of SP-CTA, but the dose can be reduced by altering the sampling interval.