American Journal of Neuroradiology, Vol 4, Issue 2 195-199, Copyright © 1983 by American Society of Neuroradiology
ARTICLES |
Optimization of xenon-enhanced CT studies: beam energy, enhancement, root mean square deviation, and repeatability
KJ Kearfott, DA Rottenberg and MD Deck
The effects of varying beam energy on the computed tomographic (CT) enhancement-to-noise (S:N) ratio were studied experimentally with the DeltaScan 2020 and GE 8800 CT scanners and a 20-cm-diameter cyclindrical Plexiglas phantom containing 11 50 ml syringes filled with varying amounts of xenon and iodine. Enhancements of 54.2, 36.7, and 31.7 Hounsfield units (H)/mg l/ml were measured with the DeltaScan 2020 at 70, 100, and 120 kVp, respectively, with corresponding root mean square deviations (RMSDs) of 12, 7, and 5 H for 400 mAs scans. For the GE 8800, enhancements of 48.3, 37.6, and 32.7 H/mg l/ml were measured at 80, 100, and 120 kVp with RMSDs of 13, 8, and 7 H for 9.6 sec 320 mA scans (3.3 msec pulse). RMSD was independent of enhancement over the range of iodine concentrations studied (0-1.5 mg l/ml) and was only a weak function of region-of-interest (ROI) size. For repeated scans with the DeltaScan 2020, measurements in 17 X 17 pixel regions were reproducible to within 0.8 H for all techniques and a drift in calibration of less than 6% was observed after 8 months of clinical use. For both the DeltaScan 2020 and the GE 8800, at the milliamperage studied, lower-energy techniques offered no advantage over 120 kVp technique for xenon CT measurements of regional cerebral blood flow, which are feasible using either of these scanners.
This article has been cited by other articles:
![]() |
T. C. Kreck, M. A. Krueger, W. A. Altemeier, S. E. Sinclair, H. T. Robertson, E. D. Shade, J. Hildebrandt, W. J. E. Lamm, D. A. Frazer, N. L. Polissar, et al. Determination of regional ventilation and perfusion in the lung using xenon and computed tomography J Appl Physiol, October 1, 2001; 91(4): 1741 - 1749. [Abstract] [Full Text] [PDF] |
||||
