AJDRAJNR - American Journal of Neuroradiology

Publication Preview: Published June 12, 2008

American Journal of Neuroradiology 2008;29:1441.

This Article
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
ajnr.A1171v1
29/8/1441    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via CrossRef
Google Scholar
Right arrow Articles by Hövener, J.-B.
Right arrow Articles by Gonen, O.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Hövener, J.-B.
Right arrow Articles by Gonen, O.

FUNCTIONAL

Whole-Brain N-Acetylaspartate MR Spectroscopic Quantification: Performance Comparison of Metabolite versus Lipid Nulling

J.-B. Hövener, D.J. Rigotti, M. Amann, S. Liu, J.S. Babb, P. Bachert, A. Gass, R.I. Grossman and O. Gonen

From the Department of Radiology (J.-B.H., D.J.R., S.L., J.S.B., R.I.G., O.G.), New York University School of Medicine, New York; German Cancer Research Center (J.-B.H., P.B.), Division of Medical Physics in Radiology, Heidelberg, Germany; and Departments of Neurology and Neuroradiology (M.A., A.G.), University Hospital Basel, Basel, Switzerland.

Please address correspondence to Oded Gonen, PhD, Department of Radiology, New York University School of Medicine, 650 First Ave, 6th Floor, New York, NY 10016; e-mail: oded.gonen{at}med.nyu.edu

BACKGROUND AND PURPOSE: Despite the prominent peak of N-acetylaspartate (NAA) in proton MR spectroscopy (1H-MR spectroscopy) of the adult brain and its almost exclusive presence in neuronal cells, the total amount of NAA, regarded as their marker, is difficult to obtain due to signal contamination from the skull lipids. This article compares the performance of 2 methods that overcome this difficulty to yield the whole-brain NAA signal, important for the assessment of the total disease load in diffuse neurologic disorders.

MATERIALS AND METHODS: The heads of 12 healthy volunteers, 3 women and 9 men, 31.0 ± 7.1 years of age, were scanned at 3T by using 2 nonlocalizing 1H-MR spectroscopy sequences: One nulls the NAA (TI = 940 ms) every second acquisition by inversion-recovery to cancel the signals of the lipids (T1 << TI) in an add-subtract scheme. The other nulls the signal of the lipids (TI = 155 ms) directly after each acquisition, requiring half as many averages for the same signal-to-noise ratio. Each sequence was repeated 3 times back-to-back on 3 occasions, and the comparison criteria were intrasubject precision (reproducibility) and total measurement duration.

RESULTS: NAA nulling is nearly twice as precise in its intrinsic back-to-back (5.8% versus 8.6%) as well as longitudinal (10.6% versus 19.7%) coefficients of variation compared with lipid nulling, but at the cost of double the acquisition time.

CONCLUSION: When speed is a more stringent requirement than precision, the new lipid-nulling sequence is a viable alternative. For precision in cross-sectional or longitudinal global NAA quantification, however, NAA nulling is still the approach of choice despite its x2 (~5 minutes) time penalty compared with the lipid-nulling approach.