Accuracy of Contrast-Enhanced MR Angiography in Predicting Angiographic Stenosis of the Internal Carotid Artery: Linear Regression Analysis
Gasser M Hathouta,b,
Michael J. Duhb and
Suzie M. El-Sadena,b
a Departments of Radiology, West Los Angeles Veterans Administration Hospital, Los Angeles, CA
b University of California, Los Angeles Medical Center, Los Angeles, CA

View larger version (12K):
[in a new window]
|
FIG 1. Scatter plot and regression line for digital subtraction angiographic measurement of stenosis versus contrast-enhanced MR angiographic measurement of stenosis. r = 0.967. The slope is very nearly 1, and the intercept is nearly 0, showing excellent linearity. DSA, digital subtraction angiography; Gd-MRA, contrast-enhanced MR angiography.
| |

View larger version (13K):
[in a new window]
|
FIG 2. Comparison of digital subtraction angiographic measurement of stenosis versus contrast-enhanced MR angiographic measurement of stenosis shows regression line with 95% mean confidence intervals, which reach a minimum of ±2.8%, showing excellent correlation between digital subtraction angiography and contrast-enhanced MR angiography in the mean. DSA, digital subtraction angiography; Gd-MRA, contrast-enhanced MR angiography.
| |

View larger version (14K):
[in a new window]
|
FIG 3. Comparison of digital subtraction angiographic measurement of stenosis versus contrast-enhanced MR angiographic measurement of stenosis shows regression line with 95% individual confidence intervals, which reach a minimum of ±13.6%. These intervals are significantly wider than the mean confidence intervals shown in Figure 2. DSA, digital subtraction angiography; Gd-MRA, contrast-enhanced MR angiography.
| |

View larger version (18K):
[in a new window]
|
FIG 4. Comparison of digital subtraction angiographic measurement of stenosis versus contrast-enhanced MR angiographic measurement of stenosis shows regression line with both 95% mean confidence intervals and 95% individual confidence intervals overlaid on the same graph to depict the difference in relative width. DSA, digital subtraction angiography; Gd-MRA, contrast-enhanced MR angiography.
| |

View larger version (15K):
[in a new window]
|
FIG 5. Comparison of digital subtraction angiographic measurement of stenosis versus ultrasound PSV shows regression line with 95% individual confidence intervals, which reach a minimum of ±27.3%. The fit is significantly less (r = 0.86) and the confidence intervals significantly wider than with contrast-enhanced MR angiography (Gd-MRA). DSA, digital subtraction angiography.
| |