American Journal of Neuroradiology 20:246-247 (2 1999)
© 1999 American Society of Neuroradiology
ARTICLE
Twinkling Artifact on Intracerebral Color Doppler Sonography
a From the Division of Diagnostic and Interventional Radiology, Geneva University Hospital, 1211 Geneva 14, Switzerland.
| Abstract |
|---|
|
|
|---|
Summary: Transcranial Doppler sonography shows potential as a noninvasive technique for long-term follow-up of treated intracranial saccular aneurysms. This technical note describes a color Doppler artifact related to microcoil architecture that might represent a potential pitfall in transcranial Doppler sonographic evaluation of aneurysmal cavity thrombosis, since it may be wrongly interpreted as residual flow or aneurysmal cavity recanalization.
| Introduction |
|---|
|
|
|---|
Transcranial Doppler sonography is a promising noninvasive technique for the detection and posttherapeutic follow-up of intracranial saccular aneurysms. This technical note describes a color Doppler artifact related to microcoil architecture that might represent a potential pitfall in transcranial Doppler sonographic evaluation of aneurysmal cavity thrombosis.
| Case Report |
|---|
|
|
|---|
Cerebral digital subtraction angiography was performed in a 45-year-old patient with subarachnoid hemorrhage proved at CT. Three saccular aneurysms were discovered: a broad-based, irregularly shaped aneurysm of the anterior communicating artery (ACoA), a 3-mm aneurysm of the right anterior choroidal artery (AChA) (Fig 1), and a 1-mm aneurysm of the right middle cerebral artery (MCA) bifurcation (Fig 1). The ACoA aneurysm was considered the source of the hemorrhage, and surgical clipping was performed; it was decided to treat the AChA aneurysm through an endovascular approach. The very small MCA aneurysm was left untreated. The AChA aneurysm was successfully embolized with a single Guglielmi detachable coil (diameter, 3 mm; length, 40 mm) (GDC-10 soft, Target Therapeutics, Fremont, CA) (Fig 2). Transcranial Doppler sonography was performed through a right temporal window before and after endovascular treatment as part of the embolization protocol adopted in our institution (Figs 3 and 4). At posttherapeutic transcranial Doppler sonography, the coil appeared as a hyperechoic structure with shadowing, and a color signal was noted as a rapidly changing mixture of red and blue behind the coil (Fig 4). No flow was noted close to or in the aneurysm, whereas a digital subtraction angiogram obtained immediately before transcranial Doppler sonography showed that the aneurysmal cavity was completely excluded from the cerebral circulation. As a complement to this clinical observation, an in vitro essay consisting of insonation of a Guglielmi detachable coil enclosed between two hydrated agar sheets (Sonar-Aid, Geistlich-Pharma, Switzerland) was performed, allowing us to reproduce the artifactual color signal (Fig 5).
|
|
|
| Discussion |
|---|
|
|
|---|
The color artifact reported herein corresponds to the color Doppler twinkling artifact described by Rahmouni et al (1). These authors showed that artifactual color appears to be generated at random, strongly reflecting a medium composed of individual reflectors. Color is not always restricted to blood flow and can be seen adjacent to areas of severe vascular stenosis (2) or in echo-free regions without flow (3). In the reported observation, no flow spectrum could be recorded in the color areas, and the spectrum was composed of close vertical bands with no outer wrapping. This artifact has been observed in the abdominal field, particularly in the detection of renal or ureteral calculi, but its potential interest in locating intracerebral coils has, to our knowledge, not been reported previously.
| Conclusion |
|---|
|
|
|---|
Transcranial Doppler sonography shows potential as a noninvasive technique for long-term follow-up of treated intracranial saccular aneurysms. The artifact described herein thus seems important to recognize, as it could be mistaken for residual flow or aneurysmal cavity recanalization.
| Footnotes |
|---|
101 Address reprint requests to Haleem G. Khan, MD.
| References |
|---|
|
|
|---|
-
Rahmouni A, Bargoin R, Herment A, et al. Color Doppler twinkling artifact in hyperechoic regions. Radiology 1996;199:269-271
[Abstract/Free Full Text] -
Middleton WD, Erikson S, Melson GL. Perivascular color artifact: pathologic significance and appearance on color Doppler US images. Radiology 1989;171:647-652
[Abstract/Free Full Text] -
Mitchell DG, Burns P, Needleman L. Color Doppler artifact in anechoic regions. J Ultrasound Med 1990;9:255-260[Abstract]
This article has been cited by other articles:
![]() |
T. Lupattelli The Yin-Yang Sign Radiology, March 1, 2006; 238(3): 1070 - 1071. [Full Text] [PDF] |
||||
![]() |
E. Ghersin, M. Soudack, and D. Gaitini Twinkling Artifact in Gallbladder Adenomyomatosis J. Ultrasound Med., February 1, 2003; 22(2): 229 - 231. [Full Text] [PDF] |
||||
![]() |
A. Kamaya, T. Tuthill, and J. M. Rubin Twinkling Artifact on Color Doppler Sonography: Dependence on Machine Parameters and Underlying Cause Am. J. Roentgenol., January 1, 2003; 180(1): 215 - 222. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Ustymowicz, J. Krejza, and Z. Mariak Twinkling Artifact in Color Doppler Imaging of the Orbit J. Ultrasound Med., May 1, 2002; 21(5): 559 - 563. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||