American Journal of Neuroradiology 25:60-62, January 2004
© 2004 American Society of Neuroradiology
Technical Note
INTERVENTIONAL
Aneurysm Packing with HydroCoil Embolic System versus Platinum Coils: Initial Clinical Experience
a From the Department of Radiology, Mayo Clinic, Rochester, MN
Address correspondence to Harry J. Cloft, MD, PhD, Department of Radiology, Mayo Clinic, 200 First Street SW, Rochester, MN 55905
| Abstract |
|---|
|
|
|---|
Summary: The HydroCoil Embolic System (HES; MicroVention, Aliso Viejo, CA) was developed to improve the efficacy of endovascular treatment of cerebral aneurysms. The HES may reduce recurrences of aneurysms by allowing for increased packing density as compared with platinum coils. We report our initial experience with the HES in the treatment of 11 patients with cerebral aneurysms and compare findings to those in size-matched control cases treated with platinum coils. The HES provided substantially improved volumetric packing of the aneurysm lumen relative to standard platinum coils (73% vs. 32%; P=.0001).
| Introduction |
|---|
|
|
|---|
A principal disadvantage of endovascular therapy of cerebral aneurysms with endovascular coils as compared with surgical clipping is aneurysm recurrence (13). Aneurysm recurrence following endovascular therapy is likely related, at least in part, to the density of packing of coils into the aneurysm, defined as volumetric percentage occlusion of the aneurysm cavity with the endovascular device (47). The HydroCoil Embolic System (HES; MicroVention, Aliso Viejo, CA) is designed with an expansile hydrogel that should theoretically fill more of the aneurysm lumen than standard platinum coils (8). We report our experience with the HES in 11 cases of cerebral aneurysms, with particular attention to the degree of aneurysm filling with embolic coils, and compare them to size-matched cases treated with platinum coils.
| Technique |
|---|
|
|
|---|
The HES is constructed as a hybrid hydrogel-platinum coil device (Fig 1). The initial diameter of these devices is 0.009 inch, and the expanded diameter is 0.027 inch. In blood, the hydrogel swells to its maximum diameter (three times the original coil diameter) in approximately 20 minutes.
|
We chose to apply the HES to patients with "medium"-size aneurysms, ranging from 3.5 to 8.5 mm in diameter. The aneurysms treated with HES were treated in a standardized manner. A single, standard, platinum framing coil (MicroVention, Aliso Viejo, CA; or Micrus Corporation, Mountain View, CA) was deposited in the aneurysm. HES coils (mean, 1.6 HES coils; range, 14 HES coils) were then deposited to fill most of the aneurysm volume. Finally, where indicated, the remainder of the aneurysm was then filled with small, soft platinum coils. Coils were added until no further coils would fit into the aneurysm without depositing in the parent vessel. Aneurysms included were located in wide variety of locations in both the anterior and posterior circulation.
A control group of 11 patients treated with standard platinum coils was assembled from our recent experience. These aneurysms were chosen to be similar in size to the aneurysms treated with HES and were located in a wide variety of locations in both the anterior and posterior circulation. These aneurysms were treated by adding coils until no further coils would fit into the aneurysm without depositing in the parent vessel.
None of cases required the use of balloon remodeling or adjunctive stent placement. Five patients in the HES group and four patients in the standard coil group had ruptured aneurysms. No clinical complications occurred in either the HES group or the standard coil group.
The aneurysm volume was calculated by assuming that the aneurysms were elliptical, by using the formula:
![]() |
![]() |
![]() |
The coil packing density was expressed by using the following formula:
![]() |
Aneurysm volumes, packing densities, numbers of coils per aneurysm, and number of coils per milliliter of aneurysm volume were compared between groups by using two-sample, two-tailed t tests (Table 1). The HES provided improved packing of the aneurysm lumen relative to standard platinum coils (72% vs. 32%; P=.0001) (Fig 2). With HES, there were trends toward decreased coils per aneurysm (3.9 coils/aneurysm vs 4.6 coils/aneurysm) and decreased coils per milliliter of aneurysm volume (55 coils/mL versus 80 coils/mL), but these trends were not statistically significant.
|
|
| Discussion |
|---|
|
|
|---|
Cerebral aneurysms treated with endovascular coil packing are only partially filled with coils, with most the intraluminal volume being filled initially with thrombus (9). Natural thrombolytic processes acting on this unstable thrombus may, in part, cause recurrence of aneurysms following endovascular coil therapy. An aneurysm filled with fewer coils and more thrombus could reasonably be expected to have a greater likelihood of recurrence. Indeed, low aneurysm coil packing density is associated with an increased risk of aneurysm recurrence, especially if less than 25% (47). The expanding hydrogel on the HES displaces clotting blood from the aneurysm lumen, resulting in a better packing density.
In our practice, we pack aneurysms tightly with coils, stopping coil deposition only when no more coils can be safely deposited. Because of the volume of hydrogel deposited with the HES, we suspected that we might use fewer coils overall. There were trends toward decreases in coils per aneurysm and coils per milliliter of aneurysm volume, but these trends did not reach statistical significance. This trend toward use of fewer coils would probably be amplified in larger aneurysms. In addition, HES coils are now available in longer lengths and larger diameters than were previously available, which might also contribute to a trend toward use of fewer coils. Because the hydrogel is soft and deformable, it may not block the deposition of further platinum coils in the same way that a purely platinum coil does. HES coils are more expensive than standard platinum coils, but, because we tended to use only one or two in each case, we do not feel that they added substantial cost to the procedure.
We do not advocate using the HES coil as the only coil to treat an aneurysm. We used standard coils to form a "basket" or "framework" for subsequent deposition of HES coils. We then deposited HES coils to fill most the aneurysm volume. In six of 11 cases, we used only a single HES coil. Because of the expansion of the gel, HES coils have a maximum of 5 minutes of repositioning time before detachment or removal is necessary. We did not experience any deployment difficulty related to this time limitation. The HES coil is stiffer than standard coils, and therefore it is often not generally used as the final or "finishing" coil. Rather, the final coil was often a small, soft standard platinum coil. Smaller and softer HES coils suitable for use as "finishing" coils are currently under development.
This is a limited study of initial experience at a single center. Although we are encouraged by the improvement in aneurysm packing with HES, the effectiveness of hydrogel in preventing aneurysm recurrence remains to be proved. The HydroCoil for Embolic Aneurysm Occlusion (HEAL) trial is a multicenter trial that is currently underway that is expected to yield information regarding aneurysm recurrence following treatment with HES. We look forward to the HEAL trial results for proof that improved aneurysm packing with HES will yield a decrease in aneurysm recurrence.
| Conclusion |
|---|
|
|
|---|
Our initial experience demonstrates that the HES allows substantially improved packing of the aneurysm lumen relative to standard platinum coils (72% vs. 32%; P=.0001). We were able to significantly improve aneurysm packing even in cases in which only a single HES was used. Future, larger series with follow-up will be necessary to demonstrate that this improvement in aneurysm filling correlates with a decreased rate of aneurysm recurrence.
| Footnotes |
|---|
Drs. Cloft and Kallmes receive research support from MicroVention. Dr. Kallmes is also a stockholders in MicroVention.
| References |
|---|
|
|
|---|
- Cognard C, Weill A, Spelle L, et al. Long-term angiographic follow-up of 169 intracranial berry aneurysms occluded with detachable coils.
Radiology1999; 212
:348
356
[Abstract/Free Full Text] - Debrun GM, Aletich VA, Kehrli P, et al. Selection of cerebral aneurysms for treatment using Guglielmi detachable coils: the preliminary University of Illinois at Chicago experience. Neurosurgery1998; 43 :1281 1295[Medline]
- Kuether TA, Nesbit GM, Barnwell SL. Clinical and angiographic outcomes, with treatment data, for patients with cerebral aneurysms treated with Guglielmi detachable coils: a single-center experience. Neurosurgery1998; 43 :1016 1025[Medline]
- Satoh K, Matsubara S, Hodoh H, Nagahiro S. Intracranial aneurysm embolization using interlocking detachable coils: correlation between volume embolization and coil compaction. Intervent Neuroradiol1997; 3 :125 129
- Tamatani S, Ito Y, Abe H, et al. Evaluation of the stability of aneurysms after embolization using detachable coils: correlation between stability of aneurysms and embolized volume of aneurysms.
AJNR Am J Neuroradiol2002; 23
:762
767
[Abstract/Free Full Text] - Uchiyama N, Kida S, Nomura M, et al. Significance of volume embolization ratio as a predictor of recanalization on endovascular treatment of cerebral aneurysms treated with Guglielmi Detachable Coils. Intervent Neuroradiol2000; 6 :59 63
- Kawanabe Y, Sadato A, Taki W, Hashimoto N. Endovascular occlusion of intracranial aneurysms with Guglielmi detachable coils: correlation between coil packing density and coil compaction. Acta Neurochir (Wien)2001; 143 :451 455[Medline]
- Kallmes DF, Fujiwara NH. New expandable hydrogel-platinum coil hybrid device for aneurysm embolization.
AJNR Am J Neuroradiol2002; 23
:1580
1588
[Abstract/Free Full Text] - Gruber A, Killer M, Bavinzski G, Richling B. Clinical and angiographic results of endosaccular coiling treatment of giant and very large intracranial aneurysms: a 7-year, single-center experience. Neurosurgery1999; 45 :793 803[Medline]
This article has been cited by other articles:
![]() |
J.B. White, C.G.M. Ken, H.J. Cloft, and D.F. Kallmes Coils in a Nutshell: A Review of Coil Physical Properties AJNR Am. J. Neuroradiol., August 1, 2008; 29(7): 1242 - 1246. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Dai, Y.H. Ding, M.A. Danielson, R. Kadirvel, G.A. Helm, D.A. Lewis, H.J. Cloft, and D.F. Kallmes Endovascular Treatment of Experimental Aneurysms with Use of Fibroblast Transfected with Replication-Deficient Adenovirus Containing Bone Morphogenetic Protein-13 Gene AJNR Am. J. Neuroradiol., April 1, 2008; 29(4): 739 - 744. [Abstract] [Full Text] [PDF] |
||||
![]() |
P.M. White, S.C. Lewis, H. Nahser, R.J. Sellar, T. Goddard, A. Gholkar, and on behalf of the HELPS Trial Collaboration HydroCoil Endovascular Aneurysm Occlusion and Packing Study (HELPS Trial): Procedural Safety and Operator-Assessed Efficacy Results AJNR Am. J. Neuroradiol., February 1, 2008; 29(2): 217 - 223. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y.H. Ding, D. Dai, R. Kadirvel, D.A. Lewis, H.J. Cloft, and D.F. Kallmes Relationship Between Aneurysm Volume and Histologic Healing after Coil Embolization in Elastase-Induced Aneurysms: A Retrospective Study AJNR Am. J. Neuroradiol., January 1, 2008; 29(1): 98 - 101. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Abruzzo, T. Tun, and A. Sambanis Efficient Transmicrocatheter Delivery of Functional Fibroblasts with a Bioengineered Collagen Gel-Platinum Microcoil Complex: Toward the Development of Endovascular Cell Therapy for Cerebral Aneurysms AJNR Am. J. Neuroradiol., September 1, 2007; 28(8): 1586 - 1593. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-J. Priebe Aneurysmal subarachnoid haemorrhage and the anaesthetist Br. J. Anaesth., July 1, 2007; 99(1): 102 - 118. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.J. Cloft and for the HEAL Investigators HydroCoil for Endovascular Aneurysm Occlusion (HEAL) Study: 3-6 Month Angiographic Follow-Up Results AJNR Am. J. Neuroradiol., January 1, 2007; 28(1): 152 - 154. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Bendszus and L. Solymosi Cerecyte Coils in the Treatment of Intracranial Aneurysms: A Preliminary Clinical Study AJNR Am. J. Neuroradiol., November 1, 2006; 27(10): 2053 - 2057. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. L. Brisman, J. K. Song, and D. W. Newell Cerebral aneurysms. N. Engl. J. Med., August 31, 2006; 355(9): 928 - 939. [Full Text] [PDF] |
||||
![]() |
The CARAT Investigators* Rates of Delayed Rebleeding From Intracranial Aneurysms Are Low After Surgical and Endovascular Treatment Stroke, June 1, 2006; 37(6): 1437 - 1442. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. C. Gaba, S. A. Ansari, S. S. Roy, F. A. Marden, M. A.G. Viana, and T. W. Malisch Embolization of Intracranial Aneurysms With Hydrogel-Coated Coils Versus Inert Platinum Coils: Effects on Packing Density, Coil Length and Quantity, Procedure Performance, Cost, Length of Hospital Stay, and Durability of Therapy Stroke, June 1, 2006; 37(6): 1443 - 1450. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. K. Wakhloo, B. B. Lieber, G. Canton, D. I. Levy, J. C. Lasheras, and C. Geindreau Changes of Intra-Aneurysmal Pressure during Coiling AJNR Am. J. Neuroradiol., March 1, 2006; 27(3): 471 - 474. [Full Text] [PDF] |
||||
![]() |
H.J. Cloft and for the HEAL Investigators HydroCoil for Endovascular Aneurysm Occlusion (HEAL) Study: Periprocedural Results. AJNR Am. J. Neuroradiol., February 1, 2006; 27(2): 289 - 292. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. O. Bescos, M. J. Slob, C. H. Slump, M. Sluzewski, and W. J. van Rooij Volume Measurement of Intracranial Aneurysms from 3D Rotational Angiography: Improvement of Accuracy by Gradient Edge Detection AJNR Am. J. Neuroradiol., November 1, 2005; 26(10): 2569 - 2572. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. A. Taschner, X. Leclerc, H. Rachdi, A. M. Barros, and J.-P. Pruvo Matrix Detachable Coils for the Endovascular Treatment of Intracranial Aneurysms: Analysis of Early Angiographic and Clinical Outcomes Stroke, October 1, 2005; 36(10): 2176 - 2180. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Weber, R. Siekmann, B. Kis, and D. Kuehne Treatment and Follow-Up of 22 Unruptured Wide-Necked Intracranial Aneurysms of the Internal Carotid Artery with Onyx HD 500 AJNR Am. J. Neuroradiol., September 1, 2005; 26(8): 1909 - 1915. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-S. Kang, M. H. Han, B. J. Kwon, O-K. Kwon, S. H. Kim, S. H. Choi, and K.-H. Chang Short-Term Outcome of Intracranial Aneurysms Treated with Polyglycolic Acid/Lactide Copolymer-Coated Coils Compared to Historical Controls Treated with Bare Platinum Coils: A Single-Center Experience AJNR Am. J. Neuroradiol., September 1, 2005; 26(8): 1921 - 1928. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. H. Ding, D. Dai, D. A. Lewis, H. J. Cloft, and D. F. Kallmes Angiographic and Histologic Analysis of Experimental Aneurysms Embolized with Platinum Coils, Matrix, and HydroCoil AJNR Am. J. Neuroradiol., August 1, 2005; 26(7): 1757 - 1763. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Lubicz, X. Leclerc, J.-Y. Gauvrit, J.-P. Lejeune, and J.-P. Pruvo Three-Dimensional Packing with Complex Orbit Coils for the Endovascular Treatment of Intracranial Aneurysms AJNR Am. J. Neuroradiol., June 1, 2005; 26(6): 1342 - 1348. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Canton, D. I. Levy, and J. C. Lasheras Changes in the Intraaneurysmal Pressure Due to HydroCoil Embolization AJNR Am. J. Neuroradiol., April 1, 2005; 26(4): 904 - 907. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Lubicz, X. Leclerc, J.-Y. Gauvrit, J.-P. Lejeune, and J.-P. Pruvo Endovascular Treatment of Intracranial Aneurysms with Matrix Coils: A Preliminary Study of Immediate Post-treatment Results AJNR Am. J. Neuroradiol., February 1, 2005; 26(2): 373 - 375. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Pelz, T. Andersson, P. Lylyk, M. Negoro, and M. Soderman Stroke Review: Advances in Interventional Neuroradiology 2004 Stroke, February 1, 2005; 36(2): 211 - 214. [Full Text] [PDF] |
||||
![]() |
D. F. Kallmes and H. J. Cloft The Use of Hydrocoil for Parent Artery Occlusion AJNR Am. J. Neuroradiol., September 1, 2004; 25(8): 1409 - 1410. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||