American Journal of Neuroradiology 27:1239-1240, June-July 2006
© 2006 American Society of Neuroradiology
Technical Note
BRAIN
Optimizing Dynamic T2* MR Imaging for Measurement of Cerebral Blood Flow Using Infusions for Cerebral Blood Volume
G.C. Newmana,b,
F.E. Hospoda,
S.B. Fainb,c and
T.D. Cookc
a Department of Neurology, University of Wisconsin School of Medicine, Madison, Wis
b Department of Radiology, University of Wisconsin School of Medicine, Madison, Wis
c Department of Medical Physics, University of Wisconsin School of Medicine, Madison, Wis
d Department of Biostatistics, University of Wisconsin School of Medicine, Madison, Wis
Address correspondence to George C. Newman, MD, PhD, Department of Neurosensory Sciences, Albert Einstein Medical Center, 5401 Old York Rd, Klein Suite #310, Philadelphia, PA 19141
 |
Abstract
|
|---|
Summary: We describe an approach to measuring cerebral blood
flow (CBF) based on independent measurements of cerebral blood
volume (CBV) and mean transit time (MTT) with calculation of
CBF by using the central volume theorem: CBF = CBV / MTT. This
permits optimization of the individual acquisitions and analyses.
In particular, measurement of CBV during contrast infusion,
rather than simultaneously with MTT from a single bolus, yields
values more consistent with those of other methods.
 |
Introduction
|
|---|
Routine measurement of cerebral perfusion by MR imaging remains
an elusive goal with qualitative publications still outnumbering
quantitative studies. Our laboratory is exploring alternatives
for calculating cerebral blood flow (CBF), applying the central
volume theorem, CBF = cerebral blood volume (CBV)/ mean transit
time (MTT), with independent measurements of CBV and MTT, which
permits individual optimization of the MR imaging acquisitions
and contrast schedules. In preliminary studies,
1 T2* contrast
imaging at 1.5T requires TR

2500 milliseconds to eliminate
T1 inflow effects for optimal CBV but TR

1500 milliseconds
for accurate fitting of bolus concentrationtime curves.
1,2 These mutually exclusive requirements and other results suggested
that measuring CBV by contrast infusion but MTT by using a bolus
would be advantageous.
3,4 In this study, 3 different strategies
of measuring CBF are compared in healthy volunteers to test
the hypothesis that optimized independent measurements of CBV
and MTT would yield perfusion values more consistent with those
from positron-emission tomography (PET), single-photon emission
tomography (SPECT), and xenon CT.
 |
Methods
|
|---|
All protocols were approved by our institutional review board.
Five healthy subjects (3 women; age range, 2125 years)
were recruited from medical center staff and students.
MR images were acquired on a GE Signa 1.5T MR imaging system by using a standard head RF coil (GE Medical Systems, Milwaukee, Wis). Each session included all 3 MR imaging sequences, a "clinical bolus," a "quantitative bolus," and an "infusion" (Table 1). All perfusion scans were acquired as multiple blocks of single-shot echo- planar images with a 22-cm field of view and 128 x 64 acquisition, reconstructed to 128 x 128.4,5 A T1-weighted fluid-attenuated inversion recovery with matching section thickness and orientation provided anatomy. Baseline image sets with at least 10 images were acquired before contrast arrival. The clinical bolus, used by University of Wisconsin Hospital neuroradiologists for many years for patient care, is oriented axially. The quantitative bolus, developed to optimize quantitation of signal intensity change for MTT measurements,1 and the infusion acquisition,4 developed specifically for CBV measurements, were oriented obliquely, with the third section parallel to the inferior surface of the medial frontal lobe, to minimize middle cerebral artery susceptibility artifact and provide a more accurate arterial input function. The total gadolinium dose for the quantitative bolus and infusion pair was 185 µmol/kg, less than a "double-dose" scan, and the total session dose remained within Food and Drug Administration guidelines. Perfusion data were analyzed by using programs written in MatLab (Mathworks, Natick, Mass). Four regions of interest each in cortical gray, deep gray, and white matter, and their associated arteries, were defined by using a semiautomated approach as described elsewhere4,5 to avoid venous contamination common with fully automated arterial input function finding procedures.
Signal intensity change was converted to relative contrast concentration, Crel, by using Equation (1):
| 1) |
 |
where
S is signal intensity and
S0 is the mean baseline signal
intensity. The mean region of interest relative concentration
was calculated for each time point from all selected pixels.
Details of our CBV methods have been published elsewhere.4 Calculation of CBV by the parametric area under the curve (AUC) method was based on Equation (2):
| 2) |
 |
where
St,0,
Sa,0,
Stissue(
t), and
Sartery(
t) were tissue or
arterial signals at baseline or time
t,

, the attenuation of
brain tissue (1.04 g/mL),
6 and
h corrects for the hematocrit
difference in large vessels and brain microvasculature (0.73).
7 Integration involved fitting of the arterial concentrationtime
curve to a gamma variate function and the tissue concentrationtime
curve to a 5-parameter lagged normal attenuation function
8 after
5-fold nonlinear interpolation (INTERP) to improve the Simplex
nonlinear fitting routines.
Calculation of CBV by the infusion (INF) method was based on Equation (3):
| 3) |
 |
where
n is the number of points measured within the plateau of the
infusion curve and
St(j) and
Sa(j) are the
jth signal intensity
measurements during that stable portion in the tissue region
of interest and artery, respectively, as described elsewhere.
4,5
All reported MTT values were calculated by the singular value decomposition deconvolution method using the bolus acquisition data.9 Note that, with this method, MTT is obtained directly from the integration of the concentration-time curve normalized by the maximum value of the curve. It is not necessary, or desirable, to extract CBV and CBF separately from the data and then calculate their ratio to obtain MTT by a reversal of the central volume theorem.2
CBF was calculated from the central volume theorem,10
| 4) |
 |
Statistical analyses were performed
with programs written with MatLab except for repeated measures
ANOVA, which was performed by using
R.11
 |
Results
|
|---|
CBV values from AUC analysis of the clinical bolus are unacceptably
large
(Table 2). Quantitative bolus acquisitions improved the
range of CBV considerably, but the values were still larger
than other techniques.
4 Only the infusion method yielded CBV
values within the expected range. CBV values from the clinical
and quantitative bolus AUC were 3.6 and 1.5 times those from
infusion, respectively (
P = .0015). Infusions also reduced subject
to subject variability dramatically
(Table 2).
MTT measured from the clinical and quantitative boluses were essentially equivalent (P = .25), though variability was less for the quantitative bolus.
Values of CBF obtained by using clinical bolus data were unreasonably large (Table 2). CBF obtained by using quantitative bolus data were closer to but consistently higher than values obtained by using other techniques. Only results calculated by using infusion CBV and quantitative bolus MTT produced CBF values similar to those available in the literature (P = .0003 for differences among the 3 methods). Mean subject infusion CBV varied from 0.025 to 0.030 mL/g, quantitative bolus MTT from 3.43 to 4.18 seconds, and CBF from 43.8 to 55.5 mL/100 g/min, very close to those obtained with PET, SPECT, and xenon CT.
 |
Discussion
|
|---|
This study demonstrates that it is possible to use dynamic contrast
T2*-weighted MR imaging to measure CBV, MTT, and CBF without
any normalization procedure. The advantages derive from the
separation of CBV and MTT measurements, which permits optimization
of each acquisition and analysis independently. Reasonable values
of CBF are obtained only when CBV is measured by infusion, supporting
earlier reports.
2,4 Infusion CBV also improves contrast-to-noise
ratio, subject to subject variability and sensitivity to pharmacologic
3 or pathologic
4 differences among subjects.
The quantitative bolus differs from the clinical bolus by using a thinner section, oblique section angulation, shorter TR and TE, larger flip angle, and lower contrast dosage injected over a slightly shorter time.1 The large CBV values resulting even from the optimized quantitative bolus, however, demonstrate the difficulty of obtaining accurate CBF by using only bolus data. Only infusion CBV provides the basis for CBF estimates that compare with literature values from other methods. The potential explanations for these quantitative differences have been discussed recently.5
It is feasible to design a single acquisition sequence to measure MTT and CBV, but the lack of programmable MR imaging infusion pumps prevents implementation at present. Studies suggest an optimal delay of 7 seconds between bolus and infusion. Furthermore, it should be possible to extract permeability data from the infusion data curve with the appropriate analysis.
 |
References
|
|---|
- Tudorica A, Newman GC, Delucia-Deranja E, et al. Quantification of gadodiamide-induced MR signal changes for hemodynamic measurements. Proc Intl Soc Mag Res Med 2000;3:2022
- Perkiö J, Aronen HJ, Kangasmaki A, et al. Evaluation of four postprocessing methods for determination of cerebral blood volume and mean transit time by dynamic susceptibility contrast imaging. Magn Reson Med 2003;47:97381
- Tudorica A, Li H, Hospod F, et al. Cerebral blood volume measurements by rapid contrast infusion and T2*-weighted echo planar MRI. Magn Reson Med 2002;47:114557[Medline]
- Newman GC, Delucia-Deranja E, Tudorica A, et al. Cerebral blood volume measurements by T2*-weighted MRI and contrast infusion. Magn Reson Med 2003;50:84455[Medline]
- Newman GC, Hospod FE, Patlak CS, et al. Experimental estimates of the constants relating signal change to contrast concentration for cerebral blood volume by dynamic T2* MRI. J Cereb Blood Flow Metab (in press; Advance online publication, November 30, 2005 doi:10.1038/sj.jcbfm.9600242)
- Axel L. Cerebral blood flow determination by rapid-sequence computed tomography. Radiology 1980;137:67986[Abstract/Free Full Text]
- Rempp KA, Brix G, Wenz F, et al. Quantification of regional cerebral blood flow and volume with dynamic susceptibility contrast-enhanced MR imaging. Radiology 1994;193:63741[Abstract/Free Full Text]
- Knudsen GM, Pettigrew KD, Patlak CS, et al. Blood-brain barrier permeability measurements by double-indicator method using intravenous injection. Am J Physiol 1994;266:H98799[Medline]
- Østergaard L, Weisskoff RM, Chesler DA, et al. High resolution measurement of cerebral blood flow using intravascular tracer bolus passages. Part I: Mathematical approach and statistical analysis. Magn Reson Med 1996;36:71525[Medline]
- Zierler KL. Theoretical basis of indicator-dilution methods for measuring flow and volume. Circ Res 1962;10:393407[Free Full Text]
- R Development Core Team R. A language and environment for statistical computing. Vienna, Austria: R Foundation for Statistical Computing; 2004
Received October 25, 2005;
accepted after revision February 22, 2006.