Technical Note
SPINE
High-Resolution Diffusion-Weighted MR Imaging of the Human Lumbosacral Plexus and Its Branches Based on a Steady-State Free Precession Imaging Technique at 3T
Z.W. Zhanga,
L.J. Songb,
Q.F. Menga,
Z.P. Lia,
B.N. Luoa,
Y.H. Yanga and
Z. Peib
a Department of Radiology, First Affiliated Hospital of Sun Yat-Sen University, Guangzhou, Guangdong, P.R. China
b Department of Neurology, First Affiliated Hospital of Sun Yat-Sen University, Guangzhou, Guangdong, P.R. China
Please address correspondence to Quanfei Meng, MD, Department of Radiology, 58# Zhongshan 2nd Rd, Guangzhou, 510080, P.R. China; e-mail: cjr.mengquanfei{at}vip
SUMMARY: 3D diffusion-weighted steady-state free precession
imaging (3D DW-SSFP) with isotropic resolution was performed
to delineate structures of the human lumbosacral plexus (LSP).
3D DW-SSFP clearly revealed detailed anatomy of the LSP and
its branches. Our data suggest that the sequence based on 3D
DW-SSFP can be used for high-resolution MR imaging of the peripheral
nervous system.
 |
Introduction
|
|---|
MR imaging evaluation of the normal peripheral nerve anatomy
and diseases is mainly dependent on 2D MR imaging techniques,
including T1-weighted spin-echo, T2-weighted fast spin-echo,
and inversion-recovery sequences with fat saturation.
1–7 Although these techniques can produce excellent-quality images,
they have limitations in describing deliberate orientations
of the targeted nerves, in that section gaps in these techniques
lead to lower anatomic coverage and less quantitative information.
Furthermore, T2-weighted spin-echo techniques cannot image smaller
nerves in the periphery because the nerves cannot be distinguished
from blood vessels on T2-weighted spin-echo images.
8 A combination
of diffusion-weighted imaging (DWI) with fat-suppressed T2-weighted
sequences has been proposed to overcome these technical limitations.
Additionally, DWI based on spin-echo with an echo-planar readout
has been used to evaluate the anatomy of the peripheral nervous
system.
9–10 However, these conventional techniques have
limited spatial resolution or low signal-to-noise ratio and
often produce severe image distortion.
8,11
It is difficult to reveal the detailed anatomy of the lumbosacral plexus (LSP) by using conventional MR imaging in that many important branches of the LSP are small, and contrast of the various anatomic structures is not sufficient to exhibit their discrete identification. The purpose of this study is to describe a high-spatial-resolution 3D diffusion-weighted steady-state free precession (3D DW-SSFP) sequence and prospectively evaluate its feasibility in human LSP imaging at 3T.
 |
Materials and Methods
|
|---|
Twenty-four healthy subjects (10 men, 14 women; mean age, 42.2
years; range, 24–69 years) underwent high-spatial-resolution
T2-weighted 3D DW-SSFP imaging with a generalized autocalibrating
partially parallel acquisition (GRAPPA) technique on a clinical
3T MR imager (Trio; Siemens Medical Solutions, Erlangen, Germany)
with 2-element circularly polarized (CP) phased-array body coils
combined with CP phased-array spinal coils.
After optimization of the sequence in 3 healthy subjects (1 woman and 2 men, 30–36 years of age), we used the following parameters: TR/TE, 14.16/4.61 ms; flip angle, 30°; FOV, 360 mm; matrix, 512 x 512; bandwidth, 184 Hz per pixel; fat suppression, water excitation technique; diffusion moments, 90 mT/m (*)msec; 3D slab thickness, 256 mm; voxel dimension, 0.8 x 0.8 x 0.8 mm (320 partitions); number of acquisitions, 1; acquisition time, 10 minutes 49 seconds. The diagram of the 3D DW-SSFP sequence used in this study is shown in Fig 1.

View larger version (8K):
[in this window]
[in a new window]
|
Fig 1. The 3D DW-SSFP sequence diagram used in the study. Diffusion gradients in the section direction were applied to reduce the signal intensity of fluid along the axis, and flow-compensation gradients in the read direction were applied to reduce motion sensitivity along this axis. DG indicates diffusion gradient; ACQ, acquisition; RF, radio-frequency; GS, section-selection gradient, Gp, phase-encoding gradient, Gr, read-out gradient.
| |
The postprocessing techniques, including maximum intensity projection and multiplanar reformation, were used to generate images of the LSP with the commercial software on the console of the MR imaging scanner. The datasets from 3D DW-SSFP sequences in all subjects were reconstructed in arbitrary planes oriented in relation to the course of the LSP and its branches. All images were analyzed by 2 experienced neuroradiologists. Each radiologist made initial evaluations independently, and any disagreements regarding final conclusions were resolved by consensus. The certainty of identifying the LSP and main branches was individually determined in each patient by using a 3-score scale system (2, identified with certainty; 1, most probable identification; 0, not identified).
The study protocol was approved by the local ethics committee and the review board of our department, and informed consent was obtained from all subjects before MR imaging examination.
 |
Results
|
|---|
In arbitrary planes, the nerve roots from L1 to S4, the spinal
nerve ganglion, and the proximal branches of the LSP were identified
(score of 1 or 2) in 48 (100%); the main branches of the lumbar
plexus including the obturator nerves and the femoral and lateral
femoral cutaneous nerves within the psoas major muscle were
identified (score of 1 or 2) in 22 (92%), 2 (8%), and 16 (67%),
respectively. The main branches of the sacral plexus including
the sciatic nerves, the posterior femoral cutaneous nerve, and
the gluteal nerves were identified (score of 1 or 2) in 24 (100%),
20 (83%), and 20 (83%), respectively. Typical imaging results
with the 3D DW-SSFP sequence are shown in
Figs 2–
4.

View larger version (80K):
[in this window]
[in a new window]
|
Fig 2. A–C, Thin-section (20 mm) coronal reconstruction MR images in a 45-year-old volunteer obtained with a 3D DW-SSFP sequence. These reconstruction planes clearly demonstrate the LSP and its components.
| |

View larger version (77K):
[in this window]
[in a new window]
|
Fig 3. A, Thin-section (20 mm) oblique sagittal reconstruction plane in a 35-year-old volunteer clearly demonstrates the components of the sacral plexus. B, Thin-section (20 mm) oblique sagittal reconstruction plane in a 28-year-old volunteer. The relationship between the nerves and the piriformis muscle is clearly defined. C, The tiny branches of LSP are shown. The white arrow indicates the superior gluteal nerve, the black arrow indicates the posterior cutaneous nerves of thigh, and the arrowhead indicates the sciatic nerve
| |

View larger version (79K):
[in this window]
[in a new window]
|
Fig 4. A–C, Curved thin-section (20 mm) coronal reconstruction MR images in a 29-year-old volunteer obtained with a 3D DW-SSFP sequence. The curved reconstruction plane (A) along the ganglion clearly demonstrates the ganglions of LSP bilaterally, the curved reconstruction plane (B) along the obturator nerve clearly demonstrates the obturator nerve (white arrow), and the curved reconstruction plane (C) along the left obturator nerve clearly demonstrates the obturator nerve (white arrow) and its branches (white arrowhead).
| |
 |
Discussion
|
|---|
The 3D DW-SSFP sequence in this study is a high-resolution diffusion-weighted
T2-weighted method of steady-state gradient-echo imaging combined
with fat-suppression and flow-compensation techniques. Our sequence
has several advantages. First, the signal intensity of the sequence
was formed by a spin-echo mechanism with characteristics of
a spin-echo sequence and thus with dominating T2 contrast as
compared with other spoiled or refocused gradient-echo techniques
because the echo of the sequence is formed by a radio-frequency
excitation that is unusual for gradient-echo sequences. Therefore,
the influence of the local magnetic field in homogeneities can
be reduced on the spin relaxation.
12
Second, the detection of small nerve structures, especially those located near large vessels, is largely improved in that the sequence can completely eliminate the signal intensity from the vessels. The SSFP signal intensity heavily depends on the steady-state condition, and vessels do not produce signal intensity because it cannot construct the steady-state condition.13 In addition, the image contrast is further improved by using the DWI technique combined with the SSFP sequence, which completely suppresses the high signal intensity of the stationary fluid, such as urine and synovia. As a result, the signal intensity of stationary fluid is very low in contrast to the high signal intensity of nerves and the intermediate signal intensity of adjacent pelvic parenchymal structures on DWI images. Furthermore, the fat signal intensity is completely suppressed, and the chemical shift effect and motion-related phase errors are reduced by using a spectral spatial water-excitation pulse. With this technique, only water is excited by using section-selective composite pulses, whereas lipid spins are left in equilibrium, thereby producing no signal intensity.14
Besides, this sequence is compatible with the GRAPPA technique at 3T. The combination of this sequence with GRAPPA allows performance of high-resolution imaging within a reasonable acquisition time, and image disturbances due to B0 field inhomogeneity in a high-field MR imager can be further decreased by the technique. Given that the thin contiguous sections and high isotropic resolution can depict tiny structures or lesions, any images can be reconstructed in arbitrary orientation by the multiplanar reconstructive technique without image degradation.
The limitations of this study include the motion, pulsation, and ghost artifacts in the lower abdomen. These artifacts can degrade or even disturb the quality of the images, especially the quality of images on the nerves close to the abdominal wall.
 |
Conclusion
|
|---|
In the present study, 3D DW-SSFP clearly revealed detailed anatomy
of the LSP and its branches. These data suggest that the sequence
based on 3D DW-SSFP can be used for high-resolution imaging
of the peripheral nervous system.
 |
References
|
|---|
- Gierada DS, Erickson SJ, Haughton VM, et al. MR imaging of the sacral plexus: normal findings. AJR Am J Roentgenol 1993;160:1059–65[Abstract/Free Full Text]
- Blake LC, Robertson WD, Hayes CE. Sacral plexus: optimal imaging planes for MR assessment. Radiology 1996;199:767–72[Abstract/Free Full Text]
- Gierada DS, Erickson SJ. MR imaging of the sacral plexus: abnormal findings. AJR Am J Roentgenol 1993;160:1067–71[Abstract/Free Full Text]
- Maravilla K, Aagaard B, Kliot M. MR neurography: MR imaging of peripheral nerves. Magn Reson Imaging Clin N Am 1998;6:179–94[Medline]
- Filler AG, Howe FA, Hayes CE, et al. Magnetic resonance neurography. Lancet 1993;341:659–61[CrossRef][Medline]
- Maravilla KR, Bowen BC. Imaging of the peripheral nervous system: evaluation of peripheral neuropathy and plexopathy. AJNR Am J Neuroradiol 1998;19:1011–23[Medline]
- Moore KR, Tsuruda JS, Dailey AT. The value of MR neurography for evaluating extraspinal neuropathic leg pain: a pictorial essay. AJNR Am J Neuroradiol 2001;22:786–94[Abstract/Free Full Text]
- Filler AG, Maravilla KR, Tsuruda JS. MR neurography and muscle MR imaging for image diagnosis of disorders affecting the peripheral nerves and musculature. Neurol Clin 2004;22:643–82[CrossRef][Medline]
- Takahara T, Imai Y, Yamashita T, et al. Diffusion weighted whole body imaging with background body signal suppression (DWIBS): technical improvement using free breathing, STIR and high resolution 3D display. Radiat Med 2004;22:275–82[Medline]
- Tsuchiya K, Fujikawa A, Tateishi H, et al. Visualization of cervical nerve roots and their distal nerve fibers by diffusion-weighted scanning using parallel imaging. Acta Radiol 2006;47:599–602[CrossRef][Medline]
- Colagrande S, Carbone SF, Carusi LM, et al. Magnetic resonance diffusion-weighted imaging: extraneurological applications. Radiol Med 2006;111:392–419. Epub 2006 Apr 11[CrossRef]
- Hoffmann KT, Hosten N, Meyer BU, et al. CSF flow studies of intracranial cysts and cyst-like lesions achieved using reversed fast imaging with steady-state precession MR sequences. AJNR Am J Neuroradiol 2000;21:493–502[Abstract/Free Full Text]
- Taupitz M, Speidel A, Hamm B, et al. T2-weighted breath-hold MR imaging of the liver at 1.5 T: results with a three-dimensional steady-state free precession sequence in 87 patients. Radiology 1995;194:439–46[Abstract/Free Full Text]
- Schick F, Forster J, Machann J, et al. Highly selective water and fat imaging applying multislice sequences without sensitivity to B1 field inhomogeneities. Magn Reson Med 1997;38:269–74[Medline]
Received July 23, 2007;
accepted after revision August 24, 2007.