PT - JOURNAL ARTICLE AU - Javier Oliván Bescós AU - Marjan J. Slob AU - Cornelis H. Slump AU - Menno Sluzewski AU - Willem Jan van Rooij TI - Volume Measurement of Intracranial Aneurysms from 3D Rotational Angiography: Improvement of Accuracy by Gradient Edge Detection DP - 2005 Nov 01 TA - American Journal of Neuroradiology PG - 2569--2572 VI - 26 IP - 10 4099 - http://www.ajnr.org/content/26/10/2569.short 4100 - http://www.ajnr.org/content/26/10/2569.full SO - Am. J. Neuroradiol.2005 Nov 01; 26 AB - Summary: Manual volume measurement of intracranial aneurysms from 3D rotational angiography varies on different threshold settings and, therefore, is operator-dependent. We developed and validated a method based on automatic gradient edge detection that is independent on threshold settings and provides an accurate and reproducible volume measurement. This method was compared with manual volume calculation in 13 aneurysm phantoms, and the results were significantly more accurate.