PT - JOURNAL ARTICLE AU - W Wichmann AU - K Von Ammon AU - U Fink AU - T Weik AU - G M Yasargil TI - Aneurysm clips made of titanium: magnetic characteristics and artifacts in MR. DP - 1997 May 01 TA - American Journal of Neuroradiology PG - 939--944 VI - 18 IP - 5 4099 - http://www.ajnr.org/content/18/5/939.short 4100 - http://www.ajnr.org/content/18/5/939.full SO - Am. J. Neuroradiol.1997 May 01; 18 AB - PURPOSE To evaluate the magnetic characteristics, artifact formation, and implant safety of titanium aneurysm clips for use in MR imaging.METHODS Aneurysm clips made of titanium alloy TiAl6V4 were tested in a magnetometer to determine their magnetic susceptibility and in a 1.5-T MR imager using both a geometric phantom and an animal model. A commercially available alpha-Phynox clip served as the reference standard.RESULTS We found minimal magnetization and a significant reduction in image artifacts with the titanium clip as compared with the Phynox clip.CONCLUSION The titanium clips improve image quality, biocompatibility, and patient safety in medical MR applications.