TY - JOUR T1 - Thrombogenicity of hydrophilic and nonhydrophilic microcatheters and guiding catheters. JF - American Journal of Neuroradiology JO - Am. J. Neuroradiol. SP - 1243 LP - 1251 VL - 18 IS - 7 AU - D F Kallmes AU - J K McGraw AU - A J Evans AU - J M Mathis AU - R W Hergenrother AU - M E Jensen AU - H J Cloft AU - M Lopes AU - J E Dion Y1 - 1997/08/01 UR - http://www.ajnr.org/content/18/7/1243.abstract N2 - PURPOSE To assess in a swine model the in vivo thrombogenicity of various microcatheters and guiding catheters as a function of catheter material, catheter coating, and duration of implantation.METHODS Microcatheters (Tracker 18 and Fastracker 18, Target Therapeutics, Fremont, Calif; Magic 1.8, Balt, Montmorency, France; and Transit, Cordis Endovascular Systems, Miami Lakes, Fla) were placed through 6F guiding catheters (Fasguide, Target Therapeutics, and Envoy, Cordis Endovascular Systems) into the common carotid arteries of swine for 30 minutes (short term), 90 minutes (medium term), and 35 days (long term). Guiding catheters were implanted for 5 hours. At the end of the implantation periods the catheters were retracted and fixed for scanning electron microscopy.RESULTS The surface of the Fastracker microcatheter was devoid of debris after both short- and medium-term implantation. The Tracker microcatheter had minimal accumulation of cellular elements whereas the Transit microcatheter showed moderate accumulation of nondeformed red blood cells. Neither the Tracker nor the Transit microcatheter showed evidence of increasing debris accumulation after medium-term implantation as compared with short-term implantation. The Magic microcatheter was coated with gross thrombus after both short- and medium-term implantation. The Fasguide guiding catheter was nearly devoid of debris, while the Envoy guiding catheter had moderate thrombus formation. Long-term implantation of the Fastracker microcatheter was well tolerated whereas that of the Transit catheter resulted in vessel occlusion.CONCLUSIONS Hydrophilic microcatheters and guiding catheters are less thrombogenic than their nonhydrophilic counterparts, but not all hydrophilic coatings are equally hypothrombogenic. Degree of thrombogenicity depends on catheter material rather than surface morphology. Medium-term implantation did not yield increasing thrombus formation relative to short-term implantation. ER -