PT - JOURNAL ARTICLE AU - G. Zaharchuk AU - E.U. Saritas AU - J.B. Andre AU - C.T. Chin AU - J. Rosenberg AU - T.J. Brosnan AU - A. Shankaranarayan AU - D.G. Nishimura AU - N.J. Fischbein TI - Reduced Field-of-View Diffusion Imaging of the Human Spinal Cord: Comparison with Conventional Single-Shot Echo-Planar Imaging AID - 10.3174/ajnr.A2418 DP - 2011 Mar 30 TA - American Journal of Neuroradiology 4099 - http://www.ajnr.org/content/early/2011/03/31/ajnr.A2418.short 4100 - http://www.ajnr.org/content/early/2011/03/31/ajnr.A2418.full AB - BACKGROUND AND PURPOSE: DWI of the spinal cord is challenging because of its small size and artifacts associated with the most commonly used clinical imaging method, SS-EPI. We evaluated the performance of rFOV spinal cord DWI and compared it with the routine fFOV SS-EPI in a clinical population. MATERIALS AND METHODS: Thirty-six clinical patients underwent 1.5T MR imaging examination that included rFOV SS-EPI DWI of the cervical spinal cord as well as 2 comparison diffusion sequences: fFOV SS-EPI DWI normalized for either image readout time (low-resolution fFOV) or spatial resolution (high-resolution fFOV). ADC maps were created and compared between the methods by using single-factor analysis of variance. Two neuroradiologists blinded to sequence type rated the 3 DWI methods, based on susceptibility artifacts, perceived spatial resolution, signal intensity–to-noise ratio, anatomic detail, and clinical utility. RESULTS: ADC values for the rFOV and both fFOV sequences were not statistically different (rFOV: 1.01 ± 0.18 × 10-3 mm2/s; low-resolution fFOV: 1.12 ± 0.22 × 10-3 mm2/s; high-resolution fFOV: 1.10 ± 0.21 × 10-3 mm2/s; F = 2.747, P > .05). The neuroradiologist reviewers rated the rFOV diffusion images superior in terms of all assessed measures (P < 0.0001). Particular improvements were noted in patients with metal hardware, degenerative disease, or both. CONCLUSIONS: rFOV DWI of the spinal cord overcomes many of the problems associated with conventional fFOV SS-EPI and is feasible in a clinical population. From a clinical standpoint, images were deemed superior to those created by using standard fFOV methods.