PT - JOURNAL ARTICLE AU - S. Soman AU - S.J. Holdsworth AU - P.D. Barnes AU - J. Rosenberg AU - J.B. Andre AU - R. Bammer AU - K.W. Yeom TI - Improved T2* Imaging without Increase in Scan Time: SWI Processing of 2D Gradient Echo AID - 10.3174/ajnr.A3595 DP - 2013 Jun 06 TA - American Journal of Neuroradiology 4099 - http://www.ajnr.org/content/early/2013/06/06/ajnr.A3595.short 4100 - http://www.ajnr.org/content/early/2013/06/06/ajnr.A3595.full AB - BACKGROUND AND PURPOSE: 2D gradient-echo imaging is sensitive to T2* lesions (hemorrhages, mineralization, and vascular lesions), and susceptibility-weighted imaging is even more sensitive, but at the cost of additional scan time (SWI: 5–10 minutes; 2D gradient-echo: 2 minutes). The long acquisition time of SWI may pose challenges in motion-prone patients. We hypothesized that 2D SWI/phase unwrapped images processed from 2D gradient-echo imaging could improve T2* lesion detection. MATERIALS AND METHODS: 2D gradient-echo brain images of 50 consecutive pediatric patients (mean age, 8 years) acquired at 3T were retrospectively processed to generate 2D SWI/phase unwrapped images. The 2D gradient-echo and 2D SWI/phase unwrapped images were compared for various imaging parameters and were scored in a blinded fashion. RESULTS: Of 50 patients, 2D gradient-echo imaging detected T2* lesions in 29 patients and had normal findings in 21 patients. 2D SWI was more sensitive than standard 2D gradient-echo imaging in detecting T2* lesions (P < .0001). 2D SWI/phase unwrapped imaging also improved delineation of normal venous structures and nonpathologic calcifications and helped distinguish calcifications from hemorrhage. A few pitfalls of 2D SWI/phase unwrapped imaging were noted, including worsened motion and dental artifacts and challenges in detecting T2* lesions adjacent to calvaria or robust deoxygenated veins. CONCLUSIONS: 2D SWI and associated phase unwrapped images processed from standard 2D gradient-echo images were more sensitive in detecting T2* lesions and delineating normal venous structures and nonpathologic mineralization, and they also helped distinguish calcification at no additional scan time. SWI processing of 2D gradient-echo images may be a useful adjunct in cases in which longer scan times of 3D SWI are difficult to implement. Abbreviations 2D SWISWI-processed 2D GREPUphase unwrapped3D SWISWI-processed 3D GREGREgradient-echo