PT - JOURNAL ARTICLE AU - K. Buch AU - B. Li AU - M.M. Qureshi AU - H. Kuno AU - S.W. Anderson AU - O. Sakai TI - Quantitative Assessment of Variation in CT Parameters on Texture Features: Pilot Study Using a Nonanatomic Phantom AID - 10.3174/ajnr.A5139 DP - 2017 May 01 TA - American Journal of Neuroradiology PG - 981--985 VI - 38 IP - 5 4099 - http://www.ajnr.org/content/38/5/981.short 4100 - http://www.ajnr.org/content/38/5/981.full SO - Am. J. Neuroradiol.2017 May 01; 38 AB - SUMMARY: Our aim was to evaluate changes in texture features based on variations in CT parameters on a phantom. Scans were performed with varying milliampere, kilovolt, section thickness, pitch, and acquisition mode. Forty-two texture features were extracted by using an in-house-developed Matlab program. Two-tailed t tests and false-detection analyses were performed with significant differences in texture features based on detector array configurations (Q values = 0.001–0.006), section thickness (Q values = 0.0002–0.001), and acquisition mode (Q values = 0.003–0.006). Variations in milliampere and kilovolt had no significant effect.GLCMgray-level co-occurrence matrixGLGMgray-level gradient matrixGLRLgray-level run lengthMDCTmultidetector row CTRPrun percentage