TY - JOUR T1 - Cervical Spinal Cord DTI Is Improved by Reduced FOV with Specific Balance between the Number of Diffusion Gradient Directions and Averages JF - American Journal of Neuroradiology JO - Am. J. Neuroradiol. SP - 2163 LP - 2170 DO - 10.3174/ajnr.A4850 VL - 37 IS - 11 AU - A. Crombe AU - N. Alberti AU - B. Hiba AU - M. Uettwiller AU - V. Dousset AU - T. Tourdias Y1 - 2016/11/01 UR - http://www.ajnr.org/content/37/11/2163.abstract N2 - BACKGROUND AND PURPOSE: Reduced-FOV DTI is promising for exploring the cervical spinal cord, but the optimal set of parameters needs to be clarified. We hypothesized that the number of excitations should be favored over the number of diffusion gradient directions regarding the strong orientation of the cord in a single rostrocaudal axis.MATERIALS AND METHODS: Fifteen healthy individuals underwent cervical spinal cord MR imaging at 3T, including an anatomic 3D-Multi-Echo Recombined Gradient Echo, high-resolution full-FOV DTI with a NEX of 3 and 20 diffusion gradient directions and 5 sets of reduced-FOV DTIs differently balanced in terms of NEX/number of diffusion gradient directions: (NEX/number of diffusion gradient directions = 3/20, 5/16, 7/12, 9/9, and 12/6). Each DTI sequence lasted 4 minutes 30 seconds, an acceptable duration, to cover C1–C4 in the axial plane. Fractional anisotropy maps and tractograms were reconstructed. Qualitatively, 2 radiologists rated the DTI sets blinded to the sequence. Quantitatively, we compared distortions, SNR, variance of fractional anisotropy values, and numbers of detected fibers.RESULTS: Qualitatively, reduced-FOV DTI sequences with a NEX of ≥5 were significantly better rated than the full-FOV DTI and the reduced-FOV DTI with low NEX (N = 3) and a high number of diffusion gradient directions (D = 20). Quantitatively, the best trade-off was reached by the reduced-FOV DTI with a NEX of 9 and 9 diffusion gradient directions, which provided significantly fewer artifacts, higher SNR on trace at b = 750 s/mm2 and an increased number of fibers tracked while maintaining similar fractional anisotropy values and dispersion.CONCLUSIONS: Optimized reduced-FOV DTI improves spinal cord imaging. The best compromise was obtained with a NEX of 9 and 9 diffusion gradient directions, which emphasizes the need for increasing the NEX at the expense of the number of diffusion gradient directions for spinal cord DTI contrary to brain DTI.CSCcervical spinal cordFAfractional anisotropyf-FOVfull-FOVMERGEMulti-Echo Recombined Gradient EchoNDGDnumber of diffusion gradient directionsr-FOVreduced-FOV ER -