TY - JOUR T1 - Can a Stent Retriever Damage the JET 7 Reperfusion Catheter? JF - American Journal of Neuroradiology JO - Am. J. Neuroradiol. DO - 10.3174/ajnr.A6804 AU - J. Pearly Ti AU - L. Yeo AU - G. Anil Y1 - 2020/10/15 UR - http://www.ajnr.org/content/early/2020/10/15/ajnr.A6804.abstract N2 - SUMMARY: We encountered 2 separate instances of damage to JET 7 reperfusion catheters when they were used in conjunction with a stent retriever during mechanical thrombectomy. On both occasions, after 1 or 2 passes with a stent retriever, we found that the distal end of the catheter was frayed and it ballooned up on flushing with saline. This mechanical failure could potentially lead to serious complications; hence, it should be shared with fellow neurointerventionalists.DACdistal access catheters ER -