RT Journal Article SR Electronic T1 Can a Stent Retriever Damage the JET 7 Reperfusion Catheter? JF American Journal of Neuroradiology JO Am. J. Neuroradiol. FD American Society of Neuroradiology SP 2317 OP 2319 DO 10.3174/ajnr.A6804 VO 41 IS 12 A1 J. Pearly Ti A1 L. Yeo A1 G. Anil YR 2020 UL http://www.ajnr.org/content/41/12/2317.abstract AB SUMMARY: We encountered 2 separate instances of damage to JET 7 reperfusion catheters when they were used in conjunction with a stent retriever during mechanical thrombectomy. On both occasions, after 1 or 2 passes with a stent retriever, we found that the distal end of the catheter was frayed and it ballooned up on flushing with saline. This mechanical failure could potentially lead to serious complications; hence, it should be shared with fellow neurointerventionalists.DACdistal access catheters