RT Journal Article SR Electronic T1 CT angiography in the examination of patients with aneurysm clips. JF American Journal of Neuroradiology JO Am. J. Neuroradiol. FD American Society of Neuroradiology SP 455 OP 457 VO 17 IS 3 A1 P T Vieco A1 E E Morin, 3rd A1 C E Gross YR 1996 UL http://www.ajnr.org/content/17/3/455.abstract AB The use of CT angiography is described in three patients for the evaluation of intracranial clips after surgery for an aneurysm. A postprocessing technique is described in which shaded-surface-display models were superimposed on maximum intensity projection CT angiograms. CT angiograms showed residual aneurysmal filling adjacent to a clip, patency of a vessel thought to be compromised by recent clip placement, and errant position of a clip, which required repeat surgery.