RT Journal Article SR Electronic T1 Femoral Artery Complications Associated with the Mynx Closure Device JF American Journal of Neuroradiology JO Am. J. Neuroradiol. FD American Society of Neuroradiology SP 1737 OP 1740 DO 10.3174/ajnr.A2153 VO 31 IS 9 A1 J.D. Fields A1 K.C. Liu A1 D.S. Lee A1 S.J. Gonda A1 A. Dogan A1 S.H. Gultekin A1 G.M. Nesbit A1 B.D. Petersen A1 S.L. Barnwell YR 2010 UL http://www.ajnr.org/content/31/9/1737.abstract AB BACKGROUND AND PURPOSE: Devices to close a femoral arteriotomy are frequently used after catheterization for interventional radiology and cardiac procedures to decrease the time to hemostasis and ambulation and, potentially, to decrease local complications. The Mynx vascular closure device uses a sealant designed to occlude the access tract, resulting in hemostasis. MATERIALS AND METHODS: We retrospectively reviewed all cases in which the Mynx device was used and for which follow-up angiography was available. A total of 146 devices were deployed in 135 patients. A follow-up vascular study visualizing the femoral artery was performed in 26 patients (27 studies). RESULTS: There were 5 (5/27, 18%) cases of intravascular Mynx sealant on follow-up vascular imaging. Three pseudoaneurysms (3/27, 11%) were identified. CONCLUSIONS: In this small study, intravascular sealant and pseudoaneurysms were found frequently after femoral arterial closure with the Mynx vascular closure device. MAUDEManufacturer and User Facility Device ExperiencePASperiodic-acid-Schiff