TABLE 3:

Greatest diameter (mm) in 3 orthogonal planes of signal loss due to coil-induced susceptibility artifact

1.5T (AP/CC/LR)3T (AP/CC/LR)
Model 1, 44.9% OcclusionModel 2, 22.4% OcclusionModel 1, 44.9% OcclusionModel 2, 22.4% Occlusion
Sequence 117.2/19.4/13.212.0/16.0/12.018.9/22.5/17.112.4/20.8/14.4
Sequence 216.5/19.4/13.211.0/16.2/10.917.8/23.0/16.214.1/20.7/14.3
Sequence 314.1/17.1/11.19.2/13.4/9.616.8/20.0/13.410.2/18.9/12.5
Sequence 411.5/16.5/10.19.5/13.9/9.714.8/18.4/12.310.0/17.4/11.8
Sequence 511.6/15.6/11.19.3/14.3/9.313.4/18.3/10.89.9/17.0/11.4
Sequence 611.7/15.7/10.89.5/14.1/8.913.7/18.8/11.810.5/16.5/11.4
Sequence 711.0/15.4/10.69.2/14.0/9.613.7/18.8/10.69.8/16.0/10.6
Sequence 811.7/14.2/10.19.6/12.6/10.512.9/13.9/10.09.4/12.7/10.1
Actual Diameters of Coil-Filled Aneurysm Models (mm) (AP/CC/LR)
Model 1: 44.9% occlusionModel 2: 22.4% occlusion
10/10/99/10/10
  • Note.—AP, anterior-posterior; CC, cranial-caudal; LR, left-right.