Table 3:

Comparison of eye lens dose savings on CT brain protocols using ATCM between titled axial-versus-helical modes and 16-section-versus-64-section MDCT

CT Brain Protocol*MDCTCARE DoseMean Eye Lens Dose (mGy)Dose Savings (%) (P value)Protocol No.
Tilted axial16+15.5681.1 (.000)1
Helical82.163
Tilted axial16+15.5667.0 (.000)1
Helical6447.185
Helical16+82.1642.6 (.000)3
6447.185
  • Note:—ATCM indicates automatic tube current modulation.

  • * Comparison was made between protocols using the paired Student t test, and statistical significance was defined at P < .05.