Table 4:

Accuracy of BSCTA in detecting cerebral aneurysms of different sizes

DiameterTP (No.)TN (No.)FP (No.)FN (No.)Sensitivity (%) (No.)Specificity (%) (No.)PPV (%) (No.)NPV (%) (No.)Accuracy (%) (No.)
>5.0 mm234500100 (23/23)100 (45/45)100 (23/23)100 (45/45)100 (68/68)
    95 % CI86–10092–10086–10092–10095–100
3.1–5.0 mm354500100 (35/35)100 (45/44)100 (35/35)100 (45/45)100 (80/80)
    95 % CI90–10092–10090–10092–10094–100
≤3.0 mm16451194 (16/17)98 (45/46)94 (16/17)98 (45/46)97 (61/63)
    95 % CI73–9989–10073–9989–10089–100
  • Note:—TP indicates true-positive; TN, true-negative; FP, false-positive; FN, false-negative; PPV, positive predictive value; NPV, negative predictive value.