Table 4:

Minimum required dose for at least sufficient qualitative image quality of the cervical spine in IR and FBP images compared with 355 mAs (IR)a

Levels C1–T1Levels C1–4
IRFBPIRFBP
Image noise score165 mAs275 mAs105 mAs150 mAs
DLP = 250 mGy × cmDLP = 412 mGy × cmDLP = 160 mGy × cmDLP = 225 mGy × cm
ED = 1.3 mSvED = 2.1 mSvED = 0.8 mSvED = 1.1 mSv
Morphologic characteristics score105 mAs105 mAs45 mAs45 mAs
DLP = 160 mGy × cmDLP = 160 mGy × cmDLP = 65 mGy × cmDLP = 65 mGy × cm
ED = 0.8 mSvED = 0.8 mSvED = 0.3 mSvED = 0.3 mSv
  • Note:—ED indicates effective dose.

  • a Data presented are the lowest applied tube current–time product at which scans showed no statistically significant differences in image quality compared with scans at the highest applied tube current–time product (355 mAs IR). Dose-length product and estimated effective dose are shown for each applied tube current–time product.