Complications | | |
Spontaneous rupture of the FRED-treated aneurysm | 0 | 2 |
Spontaneous nonaneurysmal intracranial hemorrhage | 3 | 1 |
Symptomatic ischemic stroke | | |
Minor (≤1 mo) | 8 | 5 |
Major (>1 mo) | 2 | 1 |
Stent/parent artery stenosis | 5 | 3 |
Stent/parent artery occlusion | 4 | 8 |
Permanent cranial neuropathy | 0 | 0 |
Other complications | | |
Groin hematoma | 1 | |
Dissection | 1 | |
Aneurysm growth | | 2 |
Total | 24 (4.5%) | 22 (4.1%) |
Adverse events | | |
Detachment in the hub of the microcatheter | 2 | |
Poor opening on deployment | 10 | |
Torquing of stent on deployment | 1 | |
Multiple repositioning | 2 | |
Visible gap between inner and outer stent | | 1 |
Fish mouthing of the distal stent end | | 13 |
Total | 15 (2.8%) | 14 (2.6%) |
Transient morbidity (≤90 days) | 10 (1.9%) | 7 (1.3%) |
Permanent morbidity (>90 days) | 3 (0.6%) | 1 (0.2%) |
Mortality | 4 (0.75%) | 4 (0.75%) |