Abstract
Background
Medical radiation from CT should be kept as low as reasonably achievable (ALARA), particularly in young patients.
Objective
To examine radiation dose from head CT in children in a trauma center (TC) and a regional children’s hospital (RCH).
Materials and methods
A random sample of 240 children (0–3, 4–9, 10–14 years of age) from the TC were compared with a similar cohort from the RCH. All children had undergone at least one head CT scan without contrast enhancement; data from PACS and Department of Radiology Information System were used to estimate normalized effective dose (ED). Lifetime attributable risk of cancer incidence was estimated using the Biologic Effects of Ionizing Radiation (BEIR) VII report.
Results
The mean normalized ED was significantly higher in the youngest children at the TC (2.74 mSv in those aged 0–3 years vs. 2.23 mSv in those aged 10–14 years; P<0.001) and at the RCH (2.44 mSv in those aged 0–3 years vs. 1.71 mSv in those aged 10–14 years; P<0.001). Each decreasing year of age was independently associated with a 0.06 mSv higher mean normalized ED (P<0.001). After adjusting for the age difference between the institutions, the mean normalized ED was 0.44 mSv lower at the RCH than at the TC across all ages (95% CI 0.31–0.58, P<0.001). A higher lifetime attributable risk of cancer was associated with younger age.
Conclusion
The radiation dose from head CT in children as defined by the normalized ED was highest in the youngest children and varied significantly between institutions in this bi-institutional study.
Similar content being viewed by others
References
Brenner DJ, Elliston CD, Hall EJ et al (2001) Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR 176:289–296
Slovis TL (2003) Children, computed tomography radiation dose, and the as low as reasonably achievable (ALARA) concept. Pediatrics 112:971–972
Hall EJ, Brenner DJ (2008) Cancer risks from diagnostic radiology. Br J Radiol 81:362–378
Frush DP, Donnelly LF, Rosen NS et al (2003) Computed tomography and radiation risk: what pediatric health care providers should know. Pediatrics 112:951–957
Benz MG, Benz MW (2004) Reduction of cancer risk associated with pediatric computed tomography by the development of new technologies. Pediatrics 114:205–209
Thomton FJ, Paulson EK, Yoshizumi TT et al (2003) Single- versus multi-detector row CT: comparison of radiation doses and dose profiles. Acad Radiol 10:379–385
Brenner DJ, Hall EJ (2007) Computed tomography: an increasing source of radiation exposure. N Engl J Med 357:2277–2284
Mettler FA, Thomadsen BR, Bhargavan M et al (2008) Medical radiation exposure in the U.S. in 2006: preliminary results. Health Phys 95:502–507
McCollough CH, Schueler BA (2000) Calculation of effective dose. Med Phys 27:828–837
Khursheed A, Hillier MC, Shrimpton PC et al (2002) Influence of patient age on normalized effective doses calculated for CT examinations. Br J Radiol 75:819–830
Einstein AJ, Henzlova MJ, Rajaopalan S et al (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 298:317–323
ImPACT Scan Group (2009) CT dosimetry tool. www.impactscan.org/ctdosimetry.htm Accessed 9 June 2009
Health Protection Agency (2008) NRPB-SR250: Normalised organ doses for X-ray computed tomography calculated using Monte Carlo techniques. http://www.hpa.org.uk/web/HPAweb&HPAwebStandard/HPAweb_C/1195733753330 Accessed 9 June 2009
National Research Council of the National Academies (2006) Health risks from exposure to low-levels of ionizing radiation. BEIR VII Phase 2. Chapter 12, pp 267–312, National Academies Press, Washington D.C. http://books.nap.edu/openbook.php?record_id=11340&page=R1 Accessed 9 June 2009
Linton OW, Mettler FA Jr (2003) National Council on Radiation Protection and Measurements. National conference on dose reduction in CT, with an emphasis on pediatric patients. AJR 181:321–329
Brenner DJ (2002) Estimating cancer risks from pediatric CT: going from the qualitative to the quantitative. Pediatr Radiol 32:228–231
Bongartz G, Golding SJ, Jurik AG et al (1999) European Commission. European guidelines on quality criteria for computed tomography. Report EUR 16262. European Commission, Luxembourg
Shrimpton PC, Wall BF (2000) Reference doses for paediatric computed tomography. Radiat Prot Dosim 90:249–252
Feigal DW (2001) FDA Public Health Notification: Reducing radiation risk from computed tomography for pediatric and small adult patients. Food and Drug Administration, Center for Devices and Radiological Health. http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/PublicHealthNotifications/ucm062185.htm Accessed 9 June 2009
Society of Pediatric Radiology (2002) The ALARA (as low as reasonably achievable) concept in pediatric CT intelligent dose reduction. Multidisciplinary conference organized by the Society of Pediatric Radiology. August 18–19, 2001. Pediatr Radiol 32:217–313
Goske MJ, Applegate JB, Butler PF et al (2008) The Image Gently campaign: working together to change practice. AJR 190:273–274
American College of Radiology (2008) ACR appropriateness criteria. Radiation dose assessment introduction. www.acr.org/SecondaryMainMenuCategories/quality_safety/app_criteria/RRLInformation.aspx Accessed 9 June 2009
American College of Radiology (2008) CT accreditation program requirements. www.acr.org/accreditation/computed/ct_reqs.aspx Accessed 9 June 2009
Paterson A, Frush DP, Donnelly LF et al (2001) Helical CT of the body: are settings adjusted for pediatric patients? AJR 176:297–301
Pages J, Buls N, Osteaux M et al (2003) CT dose in children: a multicentre study. Br J Radiol 76:803–811
Verdun FR, Gutierrez D, Vader JP et al (2008) CT radiation dose in children: a survey to establish age-based diagnostic reference levels in Switzerland. Eur Radiol 18:1980–1986
Huda W, Vance A (2007) Patient radiation doses from adult and pediatric CT. AJR 188:540–546
Donnelly LF (2005) Reducing radiation dose associated with pediatric CT by decreasing unnecessary examinations. AJR 184:655–657
Hall EJ (2002) Lessons we have learned from our children: cancer risks from diagnostic radiology. Pediatr Radiol 32:700–706
O’Daniel JC, Stevens DM, Cody DD et al (2005) Reducing radiation exposure from survey CT scans. AJR 185:509–515
Mulkens TH, Bellinck P, Baeyaert M et al (2005) Use of an automatic exposure control mechanism for dose optimization in multi-detector row CT examinations: clinical evaluation. Radiology 237:213–223
Kalra MK, Maher MM, Toth TL et al (2004) Strategies for CT radiation dose optimization. Radiology 230:619–628
Suess C, Chen X (2002) Dose optimization in pediatric CT: current technology and future innovations. Pediatr Radiol 32:729–734
Givner A, Gurney J, O’Connor D et al (2002) Reimaging in pediatric neurotrauma: factors associated with progression of intracranial injury. J Pediatr Surg 37:381–385
Parizel PM, VanGoethem JW, Özsarlak Ö et al (2005) New developments in the neuroradiological diagnosis of craniocerebral trauma. Eur Radiol 15:569–581
Mullins ME, Lev MH, Bove P et al (2004) Comparison of image quality between conventional and low-dose nonenhanced head CT. AJNR 25:533–538
Cohnen M, Fischer H, Hamacher J et al (2000) CT of the head by use of reduced current and kilovoltage: relationship between image quality and dose reduction. AJNR 21:1654–1660
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
King, M.A., Kanal, K.M., Relyea-Chew, A. et al. Radiation exposure from pediatric head CT: a bi-institutional study. Pediatr Radiol 39, 1059–1065 (2009). https://doi.org/10.1007/s00247-009-1327-1
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00247-009-1327-1