Quantitative assessment of an MR technique for reducing metal artifact: application to spin-echo imaging in a phantom

  • Article
  • Published:
Skeletal Radiology Aims and scope Submit manuscript

Abstract 

Objective. To quantify image artifact reduction using a new technique (MARS – metal artifact reduction sequence) in vitro.

Design. Coronal T1-weighted MR images were obtained through two metal phantoms (titanium/chromium-cobalt and stainless steel femoral prostheses) immersed in water. Comparison of artifact volume was made with images obtained using conventional and modified (MARS) T1-weighted sequences. Signal intensity values outside a range of ±40% the average signal intensity for water were considered artifact and segmented into low or high signal artifact categories. Considering the arbitrary selection of this threshold value, volumetric calculations of artifact were also evaluated at ±50%, 60%, 70%, and 80% the mean signal for water.

Results. Conventional T1-weighted images produced 87% more low signal artifact and 212% more high signal artifact compared with the MARS modified T1-weighted images of the stainless steel prosthesis. Conventional T1-weighted images of the titanium prosthesis produced 84% more low signal artifact and 211% more high signal artifact than the MARS modified sequence. The level of artifact reduction was essentially uniform for the various threshold levels tested and was greatest at ±20% the global signal intensity average for water.

Conclusion. The MARS technique reduces the volume of image signal artifact produced by stainless steel and titanium/chromium-cobalt femoral prostheses on T1-weighted spin-echo images in a tissue phantom model.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received: 11 April 2000 Revision requested: 22 May 2000 Revision received: 26 October 2000 Accepted: 27 November 2000

Rights and permissions

Reprints and permissions

About this article

Cite this article

Lee, M., Janzen, D., Munk, P. et al. Quantitative assessment of an MR technique for reducing metal artifact: application to spin-echo imaging in a phantom. Skeletal Radiol 30, 398–401 (2001). https://doi.org/10.1007/s002560100332

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s002560100332

Navigation