Abstract
Objectives
To retrospectively evaluate the image quality and radiation dose of 100-kVp scans with sinogram-affirmed iterative reconstruction (IR) for unenhanced head CT in adolescents.
Methods
Sixty-nine patients aged 12–17 years underwent head CT under 120- (n = 34) or 100-kVp (n = 35) protocols. The 120-kVp images were reconstructed with filtered back-projection (FBP), 100-kVp images with FBP (100-kVp-F) and sinogram-affirmed IR (100-kVp-S). We compared the effective dose (ED), grey–white matter (GM–WM) contrast, image noise, and contrast-to-noise ratio (CNR) between protocols in supratentorial (ST) and posterior fossa (PS). We also assessed GM–WM contrast, image noise, sharpness, artifacts, and overall image quality on a four-point scale.
Results
ED was 46% lower with 100- than 120-kVp (p < 0.001). GM–WM contrast was higher, and image noise was lower, on 100-kVp-S than 120-kVp at ST (p < 0.001). CNR of 100-kVp-S was higher than of 120-kVp (p < 0.001). GM–WM contrast of 100-kVp-S was subjectively rated as better than of 120-kVp (p < 0.001). There were no significant differences in the other criteria between 100-kVp-S and 120-kVp (p = 0.072–0.966).
Conclusions
The 100-kVp with sinogram-affirmed IR facilitated dramatic radiation reduction and better GM–WM contrast without increasing image noise in adolescent head CT.
Key points
• 100-kVp head CT provides 46% radiation dose reduction compared with 120-kVp.
• 100-kVp scanning improves subjective and objective GM–WM contrast.
• Sinogram-affirmed IR decreases head CT image noise, especially in supratentorial region.
• 100-kVp protocol with sinogram-affirmed IR is suited for adolescent head CT.
Similar content being viewed by others
Abbreviations
- AEC:
-
Automatic exposure control
- ATVS:
-
Automated tube voltage selection
- CNR:
-
Contrast-to-noise ratio
- FBP:
-
Filtered back-projection
- GM:
-
Grey matter
- IR:
-
Iterative reconstruction
- ROI:
-
Region of interest
- SAFIRE:
-
Sinogram-affirmed iterative reconstruction
- WM:
-
White matter
References
Pearce MS, Salotti JA, Little MP et al (2012) Radiation exposure from CT scans in childhood and subsequent risk of leukaemia and brain tumours: a retrospective cohort study. Lancet 380:499–505
Mathews JD, Forsythe AV, Brady Z et al (2013) Cancer risk in 680,000 people exposed to computed tomography scans in childhood or adolescence: data linkage study of 11 million Australians. BMJ 346:f2360
Sprawls P (1992) AAPM tutorial. CT image detail and noise. Radiographics 12:1041–1046
Weinstein MA, Duchesneau PM, MacIntyre WJ (1977) White and gray matter of the brain differentiated by computed tomography. Radiology 122:699–702
Kalra MK, Maher MM, Toth TL et al (2004) Strategies for CT radiation dose optimization. Radiology 230:619–628
Funama Y, Awai K, Nakayama Y et al (2005) Radiation dose reduction without degradation of low-contrast detectability at abdominal multisection CT with a low-tube voltage technique: phantom study. Radiology 237:905–910
Tanaka C, Ueguchi T, Shimosegawa E et al (2006) Effect of CT acquisition parameters in the detection of subtle hypoattenuation in acute cerebral infarction: a phantom study. AJNR Am J Neuroradiol 27:40–45
Craddock C, Chen MY, Dixon RL, Schlarb CA, Williams DW 3rd (2006) The effect of skull volume and density on differentiating gray and white matter on routine computed tomography scans of the head. J Comput Assist Tomogr 30:734–738
Yu L, Bruesewitz MR, Thomas KB, Fletcher JG, Kofler JM, McCollough CH (2011) Optimal tube potential for radiation dose reduction in pediatric CT: principles, clinical implementations, and pitfalls. Radiographics 31:835–848
Thompson JR, Triolo PJ, Moore RJ, Hinshaw DB Jr, Hasso AN (1984) CT scanning phantom for normalization of infant brain attenuation. AJNR Am J Neuroradiol 5:167–170
Rollins JD, Collins JS, Holden KR (2010) United States head circumference growth reference charts: birth to 21 years. J Pediatr 156:907–913
Yamada K, Endo S, Yoshioka S, Hatazawa J, Yamaura H, Matsuzawa T (1982) Age-related changes of the cranial bone mineral: a quantitative study with computed tomography. J Am Geriatr Soc 30:756–763
Pontana F, Pagniez J, Flohr T et al (2011) Chest computed tomography using iterative reconstruction vs filtered back projection (Part 1): Evaluation of image noise reduction in 32 patients. Eur Radiol 21:627–635
Singh S, Kalra MK, Shenoy-Bhangle AS et al (2012) Radiation dose reduction with hybrid iterative reconstruction for pediatric CT. Radiology 263:537–546
Korn A, Bender B, Fenchel M et al (2013) Sinogram affirmed iterative reconstruction in head CT: improvement of objective and subjective image quality with concomitant radiation dose reduction. Eur J Radiol 82:1431–1435
Bodelle B, Klein E, Naguib NN et al (2014) Acute intracranial hemorrhage in CT: benefits of sinogram-affirmed iterative reconstruction techniques. AJNR Am J Neuroradiol 35:445–449
Bodelle B, Wichmann JL, Scholtz JE et al (2015) Iterative reconstruction leads to increased subjective and objective image quality in Cranial CT in patients with stroke. AJR Am J Roentgenol 205:618–622
Rapalino O, Kamalian S, Kamalian S et al (2012) Cranial CT with adaptive statistical iterative reconstruction: improved image quality with concomitant radiation dose reduction. AJNR Am J Neuroradiol 33:609–615
Kilic K, Erbas G, Guryildirim M et al (2013) Quantitative and qualitative comparison of standard-dose and low-dose pediatric head computed tomography: a retrospective study assessing the effect of adaptive statistical iterative reconstruction. J Comput Assist Tomogr 37:377–381
Wallace AN, Vyhmeister R, Bagade S et al (2015) Evaluation of the use of automatic exposure control and automatic tube potential selection in low-dose cerebrospinal fluid shunt head CT. Neuroradiology 57:639–644
McKnight CD, Watcharotone K, Ibrahim M, Christodoulou E, Baer AH, Parmar HA (2014) Adaptive statistical iterative reconstruction: reducing dose while preserving image quality in the pediatric head CT examination. Pediatr Radiol 44:997–1003
Ono S, Niwa T, Yanagimachi N et al (2016) Improved image quality of helical computed tomography of the head in children by iterative reconstruction. J Neuroradiol 43:31–36
European guidelines for quality criteria for computed tomography (2000) European Commission, Luxembourg
Huda W, Ogden KM, Khorasani MR (2008) Converting dose-length product to effective dose at CT. Radiology 248:995–1003
Kurudirek M (2014) Effective atomic numbers, water and tissue equivalence properties of human tissues, tissue equivalents and dosimetric materials for total electron interaction in the energy region 10 keV-1 GeV. Appl Radiat Isot 94:1–7
Johnson TR, Krauss B, Sedlmair M et al (2007) Material differentiation by dual energy CT: initial experience. Eur Radiol 17:1510–1517
Zachrisson H, Engstrom E, Engvall J, Wigstrom L, Smedby O, Persson A (2010) Soft tissue discrimination ex vivo by dual energy computed tomography. Eur J Radiol 75:e124–e128
Wang B, Gao Z, Zou Q, Li L (2003) Quantitative diagnosis of fatty liver with dual-energy CT. An experimental study in rabbits. Acta Radiol 44:92–97
Unger E, Littlefield J, Gado M (1988) Water content and water structure in CT and MR signal changes: possible influence in detection of early stroke. AJNR Am J Neuroradiol 9:687–691
Kucinski T, Vaterlein O, Glauche V et al (2002) Correlation of apparent diffusion coefficient and computed tomography density in acute ischemic stroke. Stroke 33:1786–1791
Lewis DW, Ashwal S, Dahl G et al (2002) Practice parameter: evaluation of children and adolescents with recurrent headaches: report of the quality standards subcommittee of the American academy of neurology and the practice committee of the child neurology society. Neurology 59:490–498
Strain JD (2007) ACR appropriateness criteria on headache-child. J Am Coll Radiol 4:18–23
Hirtz D, Ashwal S, Berg A et al (2000) Practice parameter: evaluating a first nonfebrile seizure in children: report of the quality standards subcommittee of the American academy of neurology, the child neurology society, and the American epilepsy society. Neurology 55:616–623
Kuppermann N, Holmes JF, Dayan PS et al (2009) Identification of children at very low risk of clinically-important brain injuries after head trauma: a prospective cohort study. Lancet 374:1160–1170
Easter JS, Bakes K, Dhaliwal J, Miller M, Caruso E, Haukoos JS (2014) Comparison of PECARN, CATCH, and CHALICE rules for children with minor head injury: a prospective cohort study. Ann Emerg Med 64:145–152
Yu L, Leng S, Chen L, Kofler JM, Carter RE, McCollough CH (2013) Prediction of human observer performance in a 2-alternative forced choice low-contrast detection task using channelized Hotelling observer: impact of radiation dose and reconstruction algorithms. Med Phys 40:041908
Goenka AH, Herts BR, Obuchowski NA et al (2014) Effect of reduced radiation exposure and iterative reconstruction on detection of low-contrast low-attenuation lesions in an anthropomorphic liver phantom: an 18-reader study. Radiology 272:154–163
Acknowledgements
The scientific guarantor of this publication is Yasuyuki Yamashita. The authors of this manuscript declare no relationships with any companies. The authors state that this work has not received any funding. No complex statistical methods were necessary for this paper. Institutional Review Board approval was obtained. Written informed consent was waived by the Institutional Review Board. No study subjects or cohorts have been previously reported. Methodology: retrospective, case-control study, performed at one institution.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Nagayama, Y., Nakaura, T., Tsuji, A. et al. Radiation dose reduction using 100-kVp and a sinogram-affirmed iterative reconstruction algorithm in adolescent head CT: Impact on grey–white matter contrast and image noise. Eur Radiol 27, 2717–2725 (2017). https://doi.org/10.1007/s00330-016-4679-6
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00330-016-4679-6