Bi-exponential proton transverse relaxation rate (R2) image analysis using RF field intensity-weighted spin density projection: potential for R2 measurement of iron-loaded liver

https://doi.org/10.1016/S0730-725X(03)00080-8Get rights and content

Abstract

A bi-exponential proton transverse relaxation rate (R2) image analysis technique has been developed that enables the discrimination of dual compartment transverse relaxation behavior in systems with rapid transverse relaxation enhancement. The technique is particularly well suited to single spin-echo imaging studies where a limited number of images are available for analysis. The bi-exponential R2 image analysis is facilitated by estimation of the initial proton spin density signal within the region of interest weighted by the RF field intensities. The RF field intensity-weighted spin density map is computed by solving a boundary value problem presented by a high spin density, long T2 material encompassing the region for analysis. The accuracy of the bi-exponential R2 image analysis technique is demonstrated on a simulated dual compartment manganese chloride phantom system with relaxation rates and relative population densities between the two compartments similar to the bi-exponential transverse relaxation behavior expected of iron loaded liver. Results from analysis of the phantoms illustrate the potential of bi-exponential R2 image analysis with RF field intensity-weighted spin density projection for quantifying transverse relaxation enhancement as it occurs in liver iron overload.

Introduction

Magnetic resonance imaging methods for the non-invasive measurement of liver iron concentration have focused primarily on quantifying the proton transverse relaxation enhancement caused by biogenic iron oxide deposits within the liver. The intent behind these studies is to determine a calibration curve that relates the reduction in hydrogen proton signal intensity within the liver with the liver iron concentration, either through the liver signal intensity from a single image [1], [2], or from the liver transverse relaxation rate (R2) calculated from multiple spin-echo images [3], [4], [5]. Of these two approaches, signal intensity ratio (SIR) methods are inherently less sensitive than relaxation rate methods, as the SIR is effectively calculated for a single point in time on the transverse magnetisation decay curve.

The factors affecting the accuracy, sensitivity, and dynamic range of transverse relaxation rate measurements are the spin-echo measurement times at which the images are acquired, the volume of the liver over which the relaxation rates are calculated, the measured signal intensity, and the image processing scheme used to model the signal decay. Where the spin-echo times are too long to enable the measurement of sufficient signal intensity for the accurate calculation of transverse relaxation times, no correlation with liver iron concentration has been found [6], [7]. However, for shortened image acquisition times, a significant correlation between R2 (1/T2) and liver biopsy iron concentration has been observed [3], [4], [5]. The dynamic range for the quantification of liver iron levels is further improved by magnetic resonance spectroscopy (MRS) techniques [8], [9]. However, MRS techniques sacrifice imaging capability for the improved signal-to-noise ratio (SNR) of measuring a single enlarged volume element at shorter echo times. Consequently, localized volume spectroscopy provides only a single value for the liver iron concentration throughout the measurement volume. R2 imaging of the liver, on the other hand, yields information on the variation in iron concentration throughout the liver, enabling both the average iron concentration to be calculated and the variation in iron concentration to be imaged [10], [11].

A limitation of the former transverse relaxation rate measurement techniques [3], [4], [5], [8], [9], [10] is that they all employ mono-exponential analysis for determination of the transverse relaxation processes within the liver. However, given the anatomic and chemical compartmentalisation of the body’s hydrogen protons [12], [13], multi-exponential transverse relaxation is expected of the various organs and tissues of the body. For hereditary hemochromatosis patients, MRS measurements [14] have shown that transverse relaxation processes within iron loaded liver are at least bi-exponential in nature.

To improve the diagnostic capability of R2 measurements of the liver, particularly in relation to the measurement of liver iron overload, the move to bi-exponential analysis of transverse relaxation processes is warranted. Further, to enable accurate bi-exponential transverse relaxation rate analysis across multiple MR platforms, the analysis technique must accommodate both instrument and patient dependencies. The key factors to be taken into account are:

  • i)

    The change or drift in imaging gain between successive spin-echo measurements.

  • ii)

    The effects of thermal and structured noise originating from both the instrument and patient on the non-zero baseline noise distribution of the magnitude image data.

  • iii)

    The noise filtering necessary for accurate determination of the relaxation parameters.

  • iv)

    The curve fitting procedure for calculation of the relaxation parameters where a small data set is available for analysis; N ≈ 2M, where M is the number of unknown parameters and N is the number of spin-echo images available. A small number of spin-echo images will typically be available for analysis where a single spin-echo imaging methodology is used [3], [4], [10].

In this paper, an R2 image analysis technique is presented which enables the discrimination of bi-exponential transverse relaxation processes in systems with rapid transverse relaxation enhancement, where the transverse relaxation times may be shorter than the minimum echo time that can be measured. The R2 image analysis is facilitated by incorporating an estimate of the initial proton spin density signal within the subject weighted by the spatial variation in RF field intensity. The RF intensity at each point within the region of interest is computed by the solution of a boundary value problem presented by a high spin density, long T2 material encompassing the region for analysis. To evaluate the precision and accuracy of the R2 imaging technique on both mono-exponential and bi-exponential transverse relaxation systems, a series of aqueous manganese chloride (MnCl2) phantoms were studied, from which the bi-exponential system was digitally simulated. The relaxation rates and relative population densities of the two compartments in the bi-exponential system were chosen according to the bi-exponential relaxation properties reported for normal and iron-loaded animal liver [15], [16].

Section snippets

Phantom system preparation

The mono-exponential R2 phantom system consisted of a series of aqueous MnCl2 solutions prepared to concentrations of 0.100, 0.200, 0.300, 0.400, 0.500, 0.600, 0.700, 0.800, 1.000, 1.200, 1.400, 1.600, 1.800, 2.000, 2.400, 2.800 and 3.200 mM. The solutions were imaged in test tubes of 25 mm internal diameter and 25 mL in volume, arranged in alternating high and low concentrations in a cardboard box padded with plastic bags filled with lard. The lard was obtained from commercially extracted pig

Noise distribution modeling

An example of the underlying noise profile in the spin-echo image data is illustrated in Fig. 1 for a region of the background in the 6 ms spin-echo image that was free of image artifacts. The histogram is fitted with both a Rician and generalized Poisson distribution, for which the respective Pearson correlation coefficients are 0.987 and 0.994. The F -test on the reduced χ-squares ratio of the Rician fit to the noise data relative to the GPD is 2.1, which indicates that the GPD is a better

Discussion

The accurate quantification of proton transverse relaxation processes in systems where there is compartmentalisation of the transverse relaxation requires multi-exponential analysis of the transverse magnetisation decay. For example, bi-exponential modeling of transverse relaxation in tissues has been shown to distinguish between the chemical compartmentalisation of hydrogen protons, such as that between water and fat [27], [28], [29], as well as that of different anatomic compartments, such as

Conclusion

The bi-exponential proton transverse relaxation rate image analysis technique presented in this study is particularly well suited to the discrimination of dual compartment transverse relaxation behavior where a limited number of spin-echo images are available for analysis. Further, the method of RF field intensity-weighted spin density projection enables the accurate determination of bi-exponential components in instances of rapid transverse relaxation enhancement, where the transverse

Acknowledgements

We are grateful to the NH&MRC for funding (Project Grant 211947). We would like to thank Dr. Jay Ives for making imaging time available at SKG Radiology’s MRI facilities at the St. John of God Hospital, Subiaco, Western Australia. Erin Robins kindly performed the radiography.

References (41)

  • K.H Cheng

    In vivo characterization of human brain by chisquares parameter mapsmultiparameter proton T2-relaxation analysis

    Magn Reson Imaging

    (1994)
  • C Andersen et al.

    Precision, accuracy, and image plane uniformity in NMR relaxation time imaging on a 1.5 T whole-body MR imaging system

    Magn Reson Imaging

    (1994)
  • P.D Jensen et al.

    Non-invasive assessment of tissue iron overload in the liver by magnetic resonance imaging

    Br J Radiol

    (1994)
  • H.L Bonkovsky et al.

    Hepatic iron concentrationnoninvasive estimation by means of MR imaging techniques

    Radiology

    (1999)
  • J.P Kaltwasser et al.

    Non-invasive quantitation of liver iron-overload by magnetic resonance imaging

    Br J Haematol

    (1990)
  • R.J Hernandez et al.

    MR evaluation of liver iron overload

    J Comput Assist Tomogr

    (1988)
  • R.M Dixon et al.

    Assessment of hepatic iron overload in thalassemic patients by magnetic resonance spectroscopy

    Hepatology

    (1994)
  • Z.Y.J Wang et al.

    Evaluation of iron overload by single voxel MRS measurement of liver T2

    JMRI

    (2002)
  • Clark PR, Chua-anusorn W, St Pierre TG. Proton transverse relaxation rate (R2) images of liver tissue: mapping local...
  • G Brix et al.

    Evaluation of proton density by magnetic resonance imagingphantom experiments and analysis of multiple component proton transverse relaxation

    Phys Med Biol

    (1990)
  • Cited by (51)

    • A general framework for hepatic iron overload quantification using MRI

      2023, Digital Signal Processing: A Review Journal
    • Effective R<inf>2</inf> relaxation rate, derived from dual-contrast fast-spin-echo MRI, enables detection of hemisphere differences in iron level and dopamine function in Parkinson's disease and healthy individuals

      2022, Journal of Neuroscience Methods
      Citation Excerpt :

      Attempts to overcome this limitation include a dual-echo SE method (Hardy et al., 2005), where the number of echoes required to fit the observed signal to the exponential decay of the apparent transverse relaxation is minimised. This has shown satisfactory results when the echo times (TE) were selected carefully (Clark et al., 2003; St Pierre et al., 2004). Multi-echo spin-echo (MESE) may be the most common alternative to the SE, using a single scan of ∼ 10 min duration to derive R2 in the clinical setting, by generating multiple echoes within one repetition with a series of radiofrequency (RF) refocussing pulses.

    • Whole body magnetic resonance imaging (MRI)

      2014, Biomedical Imaging: Applications and Advances
    • Diffuse Liver Disease

      2014, Textbook of Gastrointestinal Radiology: Volumes 1-2, Fourth Edition
    • A short-TR single-echo spin-echo breath-hold method for assessing liver T2

      2024, Magnetic Resonance Materials in Physics, Biology and Medicine
    View all citing articles on Scopus
    View full text