Elsevier

Magnetic Resonance Imaging

Volume 26, Issue 9, November 2008, Pages 1215-1220
Magnetic Resonance Imaging

Original Contribution
A comparative study at 3 T of sequence dependence of T2 quantitation in the knee

https://doi.org/10.1016/j.mri.2008.02.017Get rights and content

Abstract

Objective

T2 mapping has been used widely in detecting cartilage degeneration in osteoarthritis. Several scanning sequences have been developed in the determination of T2 relaxation times of tissues. However, the derivation of these times may vary from sequence to sequence. This study seeks to evaluate the sequence-dependent differences in T2 quantitation of cartilage, muscle, fat and bone marrow in the knee joint at 3 T.

Methods

Three commercial phantoms and 10 healthy volunteers were studied using 3 T MR. T2 relaxation times of the phantoms, cartilage, muscle, subcutaneous fat and marrow were derived using spin echo (SE), multiecho SE (MESE), fast SE (FSE) with varying echo train length (ETL), spiral and spoiler gradient (SPGR) sequences. The differences between these times were then evaluated using Student's t test. In addition, the signal-to-noise ratio (SNR) efficiency and coefficient of variation of T2 from each sequence were calculated.

Results

The average T2 relaxation time was 36.38±5.76 ms in cartilage and 34.08±6.55 ms in muscle, ranging from 27 to 45 ms in both tissues. The times for subcutaneous fat and marrow were longer and more varying, ranging from 41 to 143 ms and from 42 to 160 ms, respectively. In FSE acquisition, relaxation time significantly increases as ETL increases (P<.05). In cartilage, the SE acquisition yields the lowest T2 values (27.52±3.10 ms), which is significantly lower than those obtained from other sequences (P<.002). T2 values obtained from spiral acquisition (38.27±6.45 ms) were higher than those obtained from MESE (34.35±5.62 ms) and SPGR acquisition (31.64±4.53 ms). These differences, however, were not significant (P>.05).

Conclusion

T2 quantification can be a valuable tool for the diagnosis of degenerative disease. Several different sequences exist to quantify the relaxation times of tissues. Sequences range in scan time, SNR efficiency, reproducibility and two- or three-dimensional mapping. However, when choosing a sequence for quantitation, it is important to realize that several factors affect the measured T2 relaxation time.

Introduction

Osteoarthritis (OA) is a degenerative disease that results in morphological and biochemical changes in tissues of the joint, including cartilage, subchondral bone and bone marrow. In cartilage, it has been shown that biochemical changes in the proteoglycan and collagen matrix can precede morphological changes that happen at later stages of the disease. Therefore, detecting biochemical changes and small intracartilaginous lesions are critical for diagnosing OA and monitoring the disease progression as well as for evaluating therapeutic procedures.

Magnetic resonance imaging (MRI) has proven to be a useful noninvasive tool in imaging joint diseases. It provides multiplanar capabilities, high spatial resolution and superior sensitivity to soft-tissue details. In addition to detecting structural changes, advanced MRI techniques have been shown to have the potential of probing biochemical changes in the tissue. In particular, spin–spin or T2 relaxation times have been correlated with increase of water contents and damage to collagen network in cartilage during OA [1], [2], [3], [4].

Although T2 quantitation has been widely used in clinical trial due to its potential correlation with cartilage biochemistry, different scan sequences and acquisition parameters can result in different measured relaxation times. In order to develop a platform from which to diagnosis disease based on relaxation time, it is essential to identify how the quantitation of T2 relaxation times of musculoskeletal tissues varies as different scanning parameters are used.

There are several factors that can affect T2 quantification. Insufficient sampling of the T2 decay curve, RF and static field inhomogeneities, stimulated echoes and T1 effects can all attribute to the incorrect quantitation of in vivo relaxation times [5]. Currently, the most commonly used T2 quantification sequences are spin echo (SE) and fast SE (FSE) sequences [6]. Several pulse sequences have been developed in the determination of T2 relaxation times in the musculoskeletal system. These sequences include the multiecho SE (MESE) sequence [7] and spiral sequence [8]. Additionally, T2 relaxation time appears to decrease slightly as the magnitude of static magnetic field increases [8], [9]. Although many clinical MR scanners operate at 1.5 T, 3 T scanners are becoming more prevalent in the clinical setting. Studies show an increase of signal-to-noise ratio (SNR) and contrast-to-noise ratio efficiencies as static magnetic field increases [8], [10], [11]. Therefore, a high-field quantitation carries with it the advantage of increased sensitivity.

Even though the differences in T2 quantification using different sequences and parameters are known, comparisons of T2 values based on different sequences are very limited. Maier et al. [7] and Mendlik et al. [12] have compared T2 quantification based on SE and several multiecho sequences at 1.5 T. To the best of our knowledge, no previous studies have systematically investigated measurement differences (T2 values, SNR and reproducibility) using sequences mentioned above (SE, FSE, MESE and spiral). In addition, with the increasing application of 3-T scanners in clinical settings, it is important to document tissue T2 relaxation times differences at this field strength.

All of the sequences mentioned are two-dimensional (2D) acquisitions. A three-dimensional (3D) acquisition may be desired, however, in T2 relaxation time quantitation. 3D imaging is free from artifacts caused by slice cross-talk. Therefore, 3D sequences can generally have a thinner slice thickness and, consequently, may provide a more accurate assessment of tissue relaxation properties.

The goal of this study was, therefore, twofold: (a) to evaluate the consistency of T2 quantification in healthy musculoskeletal tissues, including cartilage, muscle, bone marrow and subcutaneous fat, in the knee using different sequences at 3 T, including SE, FSE, MESE and spiral acquisition; (b) to evaluate a new 3D T2 quantitation method and to compare the results with currently available 2D sequences.

Section snippets

Phantoms and subjects

Three commercial cylindrical T2 phantoms (Diagnostic Sonar, Livingston, UK) with different T2 values spanning the expected relaxation times of tissues of interest were studied. Each phantom was scanned from three to four times to investigate the reproducibility and quantification accuracy for each sequence. The phantoms were repositioned between scans at different locations (center, left or right to the isocenter of the scanner).

The right knees of 10 healthy volunteers (6 males, 4 females; age

Phantom

The average CV of T2 quantification was lower than 8% in all three phantoms for all sequences, showing a good overall reproducibility, as seen in Table 1. After combining data from different sequences, the average CV was the highest (5.6%) in Phantom 3, indicating the worst reproducibility. This phantom has the longest T2 relaxation time (103 ms) and the longest T1 relaxation time (926 ms). A high CV of 13.6% was observed in the FSE sequence with ETL=16 in this phantom. After combining data

Discussion

In this study, different sequences were used to quantify T2 relaxation time at 3 T in musculoskeletal tissues, including cartilage, muscle, bone marrow and subcutaneous fat. A new 3D T2 mapping technique based on SPGR sequence was also proposed. The T2 relaxation time for each tissue or phantom was different for the different sequences (Fig. 5). However, the magnitude and the direction of sequence-dependent difference between the T2 times were consistent across all tissues. For example, SE

References (17)

  • David-VaudeyE. et al.

    T2 relaxation time measurements in osteoarthritis

    Magn Reson Imaging

    (2004)
  • KornaatP.R. et al.

    MR imaging of articular cartilage at 1.5T and 3.0T: comparison of SPGR and SSFP sequences

    Osteoarthritis Cartilage

    (2005)
  • BroderickL.S. et al.

    Severity of articular cartilage abnormality in patients with osteoarthritis: evaluation with fast spin-echo MR vs arthroscopy

    AJR Am J Roentgenol

    (1994)
  • DunnT.C. et al.

    T2 relaxation time of cartilage at MR imaging: comparison with severity of knee osteoarthritis

    Radiology

    (2004)
  • MosherT.J. et al.

    Human articular cartilage: influence of aging and early symptomatic degeneration on the spatial variation of T2 — preliminary findings at 3 T1

    Radiology

    (2000)
  • PoonC.S. et al.

    Practical T2 quantitation for clinical applications

    J Magn Reson Imaging

    (1992)
  • BartlettP.A. et al.

    T2 relaxometry of the hippocampus at 3T

    AJNR Am J Neuroradiol

    (2007)
  • MaierC.F. et al.

    T2 quantitation of articular cartilage at 1.5 T

    J Magn Reson Imaging

    (2003)
There are more references available in the full text version of this article.

Cited by (90)

  • Magnetic Resonance Imaging of the Knee: Conventional and Novel Techniques

    2020, Evidence-Based Management of Complex Knee Injuries: Restoring the Anatomy to Achieve Best Outcomes
View all citing articles on Scopus
View full text