Correlation of regional cerebral blood flow measured by stable xenon CT and perfusion MRI

J Comput Assist Tomogr. 1999 Mar-Apr;23(2):257-64. doi: 10.1097/00004728-199903000-00015.

Abstract

Purpose: The purpose of this work was to investigate the validity of perfusion MRI in comparison with stable xenon CT for evaluating regional cerebral blood flow (rCBF).

Method: The rCBF was measured by xenon CT and perfusion MRI within a 24 h interval in 10 patients (mean +/- SD age 63 +/- 10 years). For perfusion MRI, absolute values of rCBF were calculated based on the indicator dilution theory after injection of 0.1 mmol/kg of Gd-DTPA. Eight to 10 regions of interest (37 mm2) were located in the white and gray matter on the rCBF images for each of the 10 patients.

Results: The mean +/- SD values of rCBF in gray matter were 48.5 +/- 14.1 ml/100 g/min measured by xenon CT and 52.2 +/- 16.4 ml/100 g/min measured by perfusion MRI. In the white matter, the rCBF was 22.6 +/- 9.1 ml/100 g/min by xenon CT and 27.4 +/- 6.8 ml/100 g/min by perfusion MRI. There was a good correlation of rCBF values between perfusion MRI and xenon CT (Pearson correlation coefficient 0.83; p < 0.0001).

Conclusion: Comparable to xenon CT, perfusion MRI provides relatively high resolution, quantitative local rCBF information coupled to MR anatomy.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Atrophy / diagnosis
  • Atrophy / physiopathology
  • Brain / blood supply
  • Brain / diagnostic imaging*
  • Brain / pathology*
  • Carotid Artery, Internal / diagnostic imaging
  • Carotid Artery, Internal / pathology
  • Carotid Stenosis / diagnosis
  • Carotid Stenosis / physiopathology
  • Cerebrovascular Circulation*
  • Contrast Media*
  • Humans
  • Hydrocephalus / diagnosis
  • Hydrocephalus / physiopathology
  • Linear Models
  • Magnetic Resonance Imaging / instrumentation
  • Magnetic Resonance Imaging / methods*
  • Magnetic Resonance Imaging / statistics & numerical data
  • Middle Aged
  • Reproducibility of Results
  • Tomography, X-Ray Computed / instrumentation
  • Tomography, X-Ray Computed / methods*
  • Tomography, X-Ray Computed / statistics & numerical data
  • Xenon*

Substances

  • Contrast Media
  • Xenon