MR fluoroscopy-guided transthoracic fine-needle aspiration biopsy: feasibility

Radiology. 2003 Aug;228(2):589-92. doi: 10.1148/radiol.2282020640. Epub 2003 Jun 20.

Abstract

The purpose of this study was to evaluate the feasibility of using an open-configuration magnetic resonance (MR) imaging system with MR fluoroscopic guidance to perform percutaneous transthoracic fine-needle aspiration biopsy in patients with lung masses. Percutaneous transthoracic aspiration biopsies were performed with MR fluoroscopic guidance in 14 patients. The masses were 2-7 cm in diameter (mean, 4.1 cm). The needle was positioned by using a free-hand technique with MR fluoroscopic guidance. The needle tip reached the target lesion, and biopsy was performed. Analysis of the biopsy specimens facilitated a specific diagnosis in all patients. Pneumothorax was noted in two patients (14%) with chronic obstructive pulmonary disease. Study results showed that the described MR fluoroscopy-guided transthoracic biopsy technique can be used safely and successfully for lung masses. MR fluoroscopy can be used to reach the target lesion easily and accurately.