X-ray scatter correction algorithm for cone beam CT imaging

Med Phys. 2004 May;31(5):1195-202. doi: 10.1118/1.1711475.

Abstract

Developing and optimizing an x-ray scatter control and reduction technique is one of the major challenges for cone beam computed tomography (CBCT) because CBCT will be much less immune to scatter than fan-beam CT. X-ray scatter reduces image contrast, increases image noise and introduces reconstruction error into CBCT. To reduce scatter interference, a practical algorithm that is based upon the beam stop array technique and image sequence processing has been developed on a flat panel detector-based CBCT prototype scanner. This paper presents a beam stop array-based scatter correction algorithm and the evaluation results through phantom studies. The results indicate that the beam stop array-based scatter correction algorithm is practical and effective to reduce and correct x-ray scatter for a CBCT imaging task.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, U.S. Gov't, P.H.S.
  • Validation Study

MeSH terms

  • Algorithms*
  • Humans
  • Phantoms, Imaging
  • Radiographic Image Enhancement / methods*
  • Radiographic Image Interpretation, Computer-Assisted / methods*
  • Radiography, Thoracic / instrumentation
  • Radiography, Thoracic / methods*
  • Reproducibility of Results
  • Scattering, Radiation
  • Sensitivity and Specificity
  • Tomography, Spiral Computed / methods*
  • X-Rays