Diffusion-weighted imaging of the parotid gland: Influence of the choice of b-values on the apparent diffusion coefficient value

J Magn Reson Imaging. 2004 Nov;20(5):786-90. doi: 10.1002/jmri.20196.

Abstract

Purpose: To determine how the ADC value of parotid glands is influenced by the choice of b-values.

Materials and methods: In eight healthy volunteers, diffusion-weighted echo-planar imaging (DW-EPI) was performed on a 1.5 T system, with b-values (in seconds/mm2) of 0, 50, 100, 150, 200, 250, 300, 500, 750, and 1000. ADC values were calculated by two alternative methods (exponential vs. logarithmic fit) from five different sets of b-values: (A) all b-values; (B) b=0, 50, and 100; (C) b=0 and 750; (D) b=0, 500, and 1000; and (E) b=500, 750, and 1000.

Results: The mean ADC values for the different settings were (in 10(-3) mm2/second, exponential fit): (A) 0.732+/-0.019, (B) 2.074+/-0.084, (C) 0.947+/-0.020, (D) 0.890+/-0.023, and (E) 0.581+/-0.021. ADC values were significantly (P <0.001) different for all pairwise comparisons of settings (A-E) of b-values, except for A vs. D (P=0.172) and C vs. D (P=0.380). The ADC(B) was significantly higher than ADC(C) or ADC(D), which was significantly higher than ADC(E). ADC values from exponential vs. logarithmic fit (P=0.542), as well as left vs. right parotid gland (P=0.962), were indistinguishable.

Conclusion: The ADC values calculated from low b-value settings were significantly higher than those calculated from high b-value settings. These results suggest that not only true diffusion but also perfusion and saliva flow may contribute to the ADC.

MeSH terms

  • Adult
  • Analysis of Variance
  • Diffusion
  • Diffusion Magnetic Resonance Imaging / methods*
  • Echo-Planar Imaging / methods
  • Female
  • Humans
  • Image Processing, Computer-Assisted / methods
  • Male
  • Parotid Gland / anatomy & histology*
  • Prospective Studies
  • Reference Values