Retrieval of prolapsed coils during endovascular treatment of cerebral aneurysms

Neuroradiology. 2006 Apr;48(4):269-72. doi: 10.1007/s00234-006-0051-x. Epub 2006 Mar 14.

Abstract

One of the feared complications during detachable coil embolization of cerebral aneurysms is herniation of a coil loop into the parent artery. Although coil protrusion of one or two loops into the parent vessel may not cause adverse events and in some instances can be ignored, the authors believe that coil retrieval is indicated if a free end is seen pulsating along the blood flow stream to prevent migration of the entire coil mass. In one patient, a microballoon was inflated across the neck of the aneurysm during retrieval of a herniated coil to prevent further coil herniation from the aneurysm sac. We present two cases in which prolapsed coils were successfully retrieved either using a microsnare and balloon combination or a microsnare alone. This report focuses on the efficacy of the Amplatz microsnare for such retrievals and the circumstances in which a herniated coil needs to be retrieved. We report two cases in which embolization coils partially migrated into the parent artery during endovascular treatment of cerebral aneurysm and were retrieved using the Amplatz Nitinol microsnare.

Publication types

  • Case Reports

MeSH terms

  • Adult
  • Aged, 80 and over
  • Alloys
  • Cerebral Angiography
  • Device Removal / methods*
  • Embolization, Therapeutic / adverse effects*
  • Embolization, Therapeutic / instrumentation*
  • Female
  • Humans
  • Intracranial Aneurysm / diagnostic imaging
  • Intracranial Aneurysm / therapy*

Substances

  • Alloys
  • nitinol