Nonlinear regularization for per voxel estimation of magnetic susceptibility distributions from MRI field maps

IEEE Trans Med Imaging. 2010 Feb;29(2):273-81. doi: 10.1109/TMI.2009.2023787. Epub 2009 Jun 5.

Abstract

Magnetic susceptibility is an important physical property of tissues, and can be used as a contrast mechanism in magnetic resonance imaging (MRI). Recently, targeting contrast agents by conjugation with signaling molecules and labeling stem cells with contrast agents have become feasible. These contrast agents are strongly paramagnetic, and the ability to quantify magnetic susceptibility could allow accurate measurement of signaling and cell localization. Presented here is a technique to estimate arbitrary magnetic susceptibility distributions by solving an ill-posed inversion problem from field maps obtained in an MRI scanner. Two regularization strategies are considered: conventional Tikhonov regularization and a sparsity promoting nonlinear regularization using the l(1) norm. Proof of concept is demonstrated using numerical simulations, phantoms, and in a stroke model in a rat. Initial experience indicates that the nonlinear regularization better suppresses noise and streaking artifacts common in susceptibility estimation.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Algorithms
  • Animals
  • Brain / anatomy & histology
  • Computer Simulation
  • Contrast Media
  • Image Processing, Computer-Assisted / methods*
  • Magnetic Resonance Imaging / methods*
  • Male
  • Models, Theoretical*
  • Nonlinear Dynamics
  • Phantoms, Imaging
  • Rats
  • Rats, Wistar
  • Stroke

Substances

  • Contrast Media